What a 35% rise in coupling-reagent prices does to a 30 mg vial
Follow the resin, not the catalogue.
TheCompound Journal
Reporting on incretins, compounding & the peptide supply chain
Identity
The two ionisation techniques in general use produce different charge distributions, different adducts and different failure modes. Certificates almost never say which was used.
Ask a laboratory which ionisation source produced a spectrum and you learn a great deal about the spectrum before you have looked at it. Electrospray on a modern instrument will give a peptide of four thousand daltons a family of multiply charged ions and a mass accuracy in the low parts per million. A linear MALDI time-of-flight will give a single predominantly singly charged ion and an accuracy measured in hundreds of parts per million. Both can answer the question, is this the right compound. Only one of them can answer the question, is this the right compound and nothing very close to it.
It is worth being exact about what a mass spectrometer does, because the imprecision propagates. The instrument generates ions from a sample, separates them according to the ratio of their mass to their charge, and counts them at a detector. The horizontal axis of every spectrum is mass-to-charge, conventionally written m/z and expressed in thomsons or in dimensionless units depending on the vendor’s software. Nothing is weighed. Nothing is measured against a reference mass in the sense that a balance measures against a calibration weight.
What follows from this is that every molecular weight on every certificate of analysis in this market is a calculated quantity, derived from a measured m/z by assigning a charge and subtracting the mass contribution of whatever adducted to the molecule to give it that charge — usually protons, sometimes sodium, occasionally potassium or ammonium. The assignment is normally straightforward and normally correct. It is nonetheless an assignment, and when it goes wrong it goes wrong by an integer factor, which is the kind of error that produces confident nonsense rather than a plausible discrepancy.
The practical consequence for a reader is a habit: when a mass figure appears, ask what was observed and what was inferred. A report that gives both — the m/z, the charge, and the derived neutral mass — has answered the question before it was asked.
Electrospray ionisation works by pumping a solution of the analyte through a fine capillary held at a potential of a few kilovolts relative to the instrument’s entrance. The liquid emerging from the tip forms a cone and then a jet of charged droplets. As solvent evaporates the droplets shrink, the charge density on their surfaces rises, and at the point where electrostatic repulsion exceeds surface tension they fission into smaller droplets. Repeat this enough times and what is left is a bare, charged analyte ion in the gas phase.
Because the charge is acquired in solution and retained through desolvation, a peptide with several basic residues will carry several protons, and the population of ions reaching the analyser is distributed across charge states. This is the defining characteristic of electrospray spectra and the reason they look bewildering to a first-time reader: a single pure compound produces four or five prominent peaks, none of them at the molecular weight.
The distribution is not noise. It carries information about the number of accessible basic sites and about the conformational state of the molecule in solution, and it shifts predictably with mobile-phase composition and pH. It also has a practical advantage that matters for identity work: dividing the mass by three or four brings a large peptide into the range where instruments achieve their best resolving power and accuracy.1
A D-amino acid substitution changes the molecule, changes its biology, and changes its mass by exactly nothing.
On what mass spectrometry is structurally unable to detectMatrix-assisted laser desorption ionisation takes a different route. The peptide is mixed with a large molar excess of a small organic compound that absorbs strongly at the laser wavelength — α-cyano-4-hydroxycinnamic acid and sinapinic acid are the usual choices for peptides and proteins respectively — and the mixture is dried on a metal target. A pulsed ultraviolet laser strikes the crystal, the matrix absorbs the energy, and a plume of matrix and analyte is ejected into the vacuum with the analyte largely intact and mostly singly protonated.
Two consequences follow. First, MALDI spectra are simple: one predominant ion per compound, at the molecular weight plus one proton, which makes them easy to read and easy to print on a certificate. Second, MALDI is markedly more tolerant of salts, buffers and heterogeneous samples than electrospray, which is why it survives in routine synthesis monitoring where electrospray would require a chromatographic clean-up first.
The trade-offs are equally real. The matrix produces intense chemical background below roughly a thousand daltons, which obscures small fragments. Ion yield varies between compounds and between spots on the same target, making MALDI a poor quantitative technique. And the achievable mass accuracy on a linear instrument at peptide molecular weights is measured in hundreds of parts per million unless a reflectron and delayed extraction are in use.2
| Compound | Monoisotopic (Da) | Average (Da) | Difference (Da) | Difference (ppm) |
|---|---|---|---|---|
| BPC-157 | 1418.68 | 1419.53 | 0.85 | 600 |
| Liraglutide | 3748.05 | 3751.20 | 3.15 | 840 |
| Semaglutide | 4111.12 | 4113.58 | 2.46 | 598 |
| Tirzepatide | 4810.47 | 4813.45 | 2.98 | 619 |
| Retatrutide | 4728.42 | 4731.30 | 2.88 | 609 |
| Tesamorelin | 5131.63 | 5135.90 | 4.27 | 831 |
| Values calculated from published molecular formulae for the free-base forms and rounded to two decimal places; salt forms and acylation variants shift these figures. The final column shows why a certificate that does not state its convention cannot be checked: the convention difference alone exceeds any plausible acceptance tolerance. | ||||
The choice between the two techniques is not a matter of quality but of question. A synthesis chemist watching a coupling proceed wants a fast, salt-tolerant check that the chain has grown by the expected residue, and MALDI on a bench instrument answers that in minutes. An analytical laboratory asked whether a submitted vial contains the labelled compound and nothing closely related to it needs the resolving power and the accuracy that electrospray into a high-field analyser provides, coupled to a chromatographic separation so that species which co-elute can at least be assigned to retention times.
Both appear in this market, and reports rarely distinguish them. That matters because the two techniques have different blind spots. MALDI can induce loss of labile modifications during desorption, so a phosphorylated or otherwise fragile species may be under-represented. Electrospray suppresses ionisation of some analytes in the presence of others, so a minor component of a mixture may be absent from a spectrum in which it is genuinely present.
A certificate stating the source therefore tells a reader which class of error to consider. The Journal has stopped asking suppliers for more testing and started asking them for this line instead, on the grounds that it costs nothing and changes what the existing test can be said to support.
This publication applies one rule to every identity claim it reports, and it is worth stating in isolation because it governs the rest. A mass measurement supports a statement about composition. Only a fragmentation or mapping experiment supports a statement about sequence. Where a source says identity was confirmed, we report that a mass was measured, unless we have seen evidence of the second kind.
The rule has consequences we accept. It makes our coverage read as more sceptical than the underlying documents, because the documents claim more than they establish. It occasionally irritates laboratories which have in fact done sequence-level work and have simply not printed it, and the remedy there is a two-line email which we are glad to receive. And it means we cannot describe any research-grade vial in this market as sequence-confirmed, because on the evidence available to us almost none are.
What the rule is not is an accusation. Nothing in this article suggests that vendors are selling material other than what they label, and the Journal has no evidence of that in respect of any company it covers. The claim is narrower and, we think, harder to argue with: the documentation in general circulation does not have the discriminating power that the language on it implies, and the gap between the two is where every avoidable dispute in this market begins.
The next piece in this department takes the document rather than the instrument as its subject: what a certificate of analysis contains, what it systematically omits, and how to check one in the time it takes to drink a coffee. Identity is one line on that page, and by the standards of the rest of it, one of the better-behaved ones.
Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.
On your point about D-amino acids: chiral amino-acid analysis after hydrolysis is not exotic and several contract laboratories offer it. The obstacle is that hydrolysis itself racemises a few per cent of most residues, so the method has a blank problem, and interpreting a low-level D content is genuinely difficult rather than merely expensive.
— V. Bhattarai, Kathmandu
An important qualification and we are glad to have it. The article implied the barrier was commercial when a substantial part of it is methodological. Recorded, and the section has been rewritten accordingly.
You state that fourteen of twenty suppliers report an MS identity test. Does that count reports supplied to you on request, or only what appears on the certificate a customer receives?
— N. Prasetyo, Surabaya
The former, which the table note now says explicitly. The count for what appears on a customer-facing certificate is lower in at least four cases, and we should have separated the two columns rather than merging them.
Follow the resin, not the catalogue.
A market that competes on one measurement will optimise that measurement. There are legitimate ways to raise a purity figure that involve no change to the product at all.
Reported from the analysis, not from a warning notice.
A reminder that a purity figure is the output of a method, and that methods differ.
Every certificate circulating in this market answers a question about molecules. Almost none answers a question about organisms, pyrogens, or the integrity of the seal.
The supplier has not disputed the finding. It has not explained the gap either.