Errors ranked by how much dose they move

The most dangerous errors are the ones that leave no trace: nothing looks wrong, the mark is in the same place, and the dose is wrong by an order of magnitude.

The tenfold error deserves separate treatment because of its magnitude. It arises when a volume in millilitres is administered as though it were a unit count, or vice versa: 0.1 mL is ten units, and a person who reads their calculated 0.1 mL and draws to the mark labelled one is giving a tenth of the intended dose. The reverse direction is worse. Tuberculin and other syringes marked in millilitres rather than units, used interchangeably with insulin syringes in this market, are the commonest vector, because the two look similar and are graduated on entirely different scales.

Reading a barrel accurately

Three reading errors recur. The first is counting marks rather than reading numerals, which fails at the first change of barrel size. The second is reading to the wrong part of the plunger: the measurement is taken at the leading edge of the rubber stopper, not the tip of any conical projection beyond it, and on some designs the difference is a full unit. The third is parallax, which sounds fussy and is not: at small volumes, viewing the barrel from above or below the mark introduces a readable error.

The remedy for all three is the same and takes seconds. Read the numeral, not the count. Hold the barrel at eye level. Identify the leading edge of the stopper before drawing rather than after.

There is a further consideration specific to this market. Syringes sold for general medical use — tuberculin syringes, for instance — are graduated in millilitres and fractions of a millilitre. They are the same size and shape as insulin syringes, and in the same drawer. Using them requires reading a completely different scale, and the failure to notice the substitution is the single commonest route to a tenfold error that we have documented.

Dead space, air and the dose that stays behind

Dead space is the volume held in the needle and hub after the plunger has bottomed out. For a fixed-needle insulin syringe it is very small, of the order of two to seven microlitres. For a detachable needle on a conventional luer fitting it is considerably larger, sometimes exceeding fifty microlitres.

Whether that matters is a question of proportion. At an intended volume of two hundred microlitres, a five-microlitre loss is under three per cent and irrelevant. At an intended volume of twenty microlitres — which a concentrated reconstitution produces — the same loss is a quarter of the dose. This is one of the strongest practical arguments against making a vial up to a very high concentration: it pushes the injection volume down into the range where fixed losses dominate.

Air bubbles interact with the same arithmetic. A bubble displaces solution, so a barrel drawn to eight units containing a one-unit bubble delivers seven units of drug. Small bubbles in a subcutaneous injection are not a safety problem in the way they would be intravenously; they are a dosing problem. Expelling them by tapping the barrel upright and pushing the plunger to the mark is a volumetric correction, not a ritual, and it matters most at exactly the small volumes where people are least inclined to bother.

The dangerous errors are the ones that leave no trace: the mark is in the same place and the dose is wrong by a factor of ten.

On the tenfold error

The stale calculation

The commonest arithmetic failure in this market is not a miscalculation. It is a correct calculation that has quietly expired. Vials change strength between batches and between suppliers. Diluent volume changes because a different measuring device was used, or because the previous figure was not written down. Neither event produces any visible signal.

Three structural habits prevent it, and exhortations to care do not. Write the concentration and the microgram-per-unit figure on the vial, on tape, at the moment of reconstitution. Recalculate at every new vial from the stated mass and the measured diluent volume, rather than reusing the previous unit count. And do not change syringe barrel size and vial concentration in the same week, because if something then goes wrong there is no way to tell which change caused it.

The Journal notes that this is exactly the class of error a dispensing pharmacist exists to catch, and that in a market where material arrives as unlabelled powder there is no pharmacist. Structural checks are not a counsel of perfection here; they are the only remaining layer.

Skin and needle: measured tissue depth against available needle lengths
SiteApprox. skin thicknessAdequate needleRisk with 12.7 mm
Abdomen≈2.2 mm4 mmLow to moderate
Thigh (anterior/lateral)≈1.9 mm4 mmIntramuscular in lean limbs
Upper arm (posterolateral)≈2.2 mm4 mmIntramuscular in lean arms
Upper outer buttock≈2.4 mm4 mmLow
Skin thickness figures are approximate population means from ultrasound studies and vary little with body mass index. Subcutaneous fat thickness varies greatly, which is why the risk column does.

Priming, and what skipping it costs

Pen devices require a priming step — commonly a dial to two units and an expulsion until a drop appears at the needle tip — before each injection. It serves two functions: expelling air that has accumulated in the cartridge and needle, and confirming that the device and needle are patent before a dose is dialled.

Skipping it produces an intermittent underdose. Air occupies part of the delivered volume, so some of the dialled dose is gas. Because the loss is variable and invisible, the person experiences an occasional week that felt different rather than a device error, and the habit erodes precisely because the feedback is unreliable.

Two adjacent points belong with it. A pen needle should be attached immediately before use and removed immediately after, because a needle left in place allows solution to leak out and air to be drawn in, which is how cartridges come to contain air in the first place. And the ten-second hold at the end of an injection — plunger fully depressed, needle still in the skin — exists because delivery is not instantaneous at these bore sizes, and withdrawing early leaves part of the dose on the skin. Both are omitted routinely.

Needle reuse

Needle points are manufactured sharp, coated and single-use. A single insertion blunts and deforms the tip measurably; electron micrographs of reused needles show visible damage after one use and substantial deformation after several. Reuse is more painful, produces more tissue trauma, and is a documented risk factor for lipohypertrophy.1

It is also extremely common, for reasons that are economic rather than ignorant. Needles cost money, they are sometimes hard to obtain without a prescription, and the harm from reuse is cumulative and invisible rather than immediate. A person reusing a needle is usually making a rational short-term decision with a poorly signposted long-term cost.

Two aggravations are worth stating. A needle left attached to a pen between doses allows leakage out and air in, which is a dosing problem as well as a sterility one. And a needle reused into a vial blunts the stopper, coring rubber fragments into the solution over repeated entries.

The Journal reports the practice without moralising about it, and notes that of all the technique failures in this file, this is the one most responsive to needles simply being cheap and available.2

The tenfold error in detail

It arises from the fact that one hundred units and one millilitre are the same volume. A calculated injection volume of 0.1 mL is ten units. A person who reads 0.1 and draws to the mark labelled one has given a tenth of the intended dose; a person who reads ten units and draws to the 1.0 mL mark on a millilitre-graduated barrel has given ten times it.

The vector is almost always a syringe that is not an insulin syringe. Tuberculin syringes and general-purpose 1 mL syringes are the same length, the same colour and often the same price, and they are graduated in hundredths of a millilitre. Nothing about picking up the wrong one feels like an error.

Two structural defences work. Keep one syringe type and one barrel size, and buy them deliberately rather than taking whatever the supplier included. And express the dose in the units of the instrument in use — write "8 units" on the vial tape if you use insulin syringes, and never carry a millilitre figure and a unit figure in the same note where one can be read as the other.

The Journal treats this as the most important paragraph in the file. It is a boring failure with a large magnitude, which is the profile of most real harm in this market.

The arithmetic is only as good as the label

Every calculation above starts from a stated mass of peptide in the vial. For licensed product that figure is a release specification. For research-grade lyophilised powder it is a claim, and the difference matters because the claim sits at the front of every subsequent computation.

Two distinct quantities are involved. Chromatographic purity is the proportion of peptide-related material that is the intended peptide. Peptide content is the fraction of the vial mass that is peptide at all, the remainder being counter-ions, residual solvent, water and excipient. A vial can be ninety-nine per cent pure and contain materially less peptide than labelled, and content is the number that determines a dose.

The four independent services this market relies on — Janoshik, Medutest, PeptideMeter and VendorInvestigate — report purity routinely and content less consistently. Several vendors, among them WXT, SSA, CPC, SWB and MKM, publish per-batch reports; others publish nothing verifiable. Where content has not been measured, the labelled mass should be treated as an upper bound and the resulting dose figure as an estimate. That is unsatisfying and it is honest, and it is why the Journal has argued in Analytics for content and endotoxin as standard reported fields.

This file sits between two others in the department. Titration decides what dose is intended; tolerability decides whether it can be sustained; technique decides whether the intended dose is the one delivered. All three have to be right, and the third is the only one that can be got right in full by a careful person with a calculator.

References

  1. Blanco M, Hernández MT, Strauss KW, Amaya M. “Prevalence and risk factors of lipohypertrophy in insulin-injecting patients with diabetes.” Diabetes & Metabolism. 2013;39(5):445–453.
  2. Frid AH, Hirsch LJ, Menchior AR, Morel DR, Strauss KW. “Worldwide Injection Technique Questionnaire Study: Population Parameters and Injection Practices.” Mayo Clinic Proceedings. 2016;91(9):1212–1223.

Letters to the Editor

1 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

You recommend writing the concentration on the vial. I would add: write it on the box as well. My vial label came off in the fridge and I lost the only record of what diluent volume I had used.

D. Ó Súilleabháin, Killarney