Nought point nine eight of a dalton: the modification that hides in plain sight

What would be needed to catch each of these, and what it would cost.

The useful way to think about a mass spectrometer is as an instrument that answers one question extremely well and is silent on several adjacent ones. Deamidation of an asparagine residue converts an amide to a carboxylic acid and adds nine hundred and eighty-four thousandths of a dalton. Oxidation of a methionine adds one oxygen atom, fifteen point nine nine five daltons. Inversion of a single stereocentre from the L to the D configuration adds nothing whatever. Those three numbers — roughly one, roughly sixteen, and zero — describe the three most consequential things that go wrong with a synthetic peptide, and an instrument’s ability to detect them falls off in exactly that order.

The instrument classes, and what each can support

A single quadrupole mass filter provides unit resolution and mass accuracy of a few tenths of a dalton. It is entirely adequate to confirm that a sample is broadly the compound expected and to detect large modifications, and it is the analyser in most low-cost LC-MS systems. It cannot resolve an isotopic envelope at peptide molecular weights and therefore cannot assign charge from spacing.

Time-of-flight analysers separate ions by the time they take to traverse a flight tube. A linear tube gives modest resolving power; adding a reflectron and delayed extraction raises it into the tens of thousands, and modern quadrupole time-of-flight hybrids achieve low single-figure parts-per-million accuracy with routine calibration. Orbital trapping instruments measure the frequency of ion oscillation in an electrostatic field and convert it by Fourier transform, delivering resolving powers from sixty thousand to several hundred thousand and sub-part-per-million accuracy with internal calibration. Fourier-transform ion cyclotron resonance remains the highest-performing class and the least common outside academic facilities.

What this hierarchy means for a reader of certificates is that the instrument named on the document sets a ceiling on what the document can claim, independent of the laboratory’s competence. An unnamed instrument leaves that ceiling unknown, which is why the Journal now treats the absence of an instrument name as a material omission rather than a stylistic one.

Isobaric and near-isobaric substitutions

Two species are isobaric if their masses are identical to the precision of the measurement, and the term covers two quite different situations. True isobars have identical elemental compositions: leucine and isoleucine are structural isomers of one another, as are the aspartate and isoaspartate products of deamidation, and no mass measurement at any resolving power will separate them. Near-isobars have different compositions that happen to give similar masses, and these are resolvable given sufficient performance.

The canonical near-isobaric pair in peptide work is glutamine against lysine, differing by 0.036 daltons — nine parts per million on a four-thousand-dalton peptide, and therefore a discrimination that requires an orbital trap or better. A second is the classic composition ambiguity in which a combination of light elements substitutes for a heavier one at nearly the same nominal mass; the mass defect of hydrogen relative to the heavier elements is what makes these separable at high resolving power and indistinguishable at low.1

The reason this matters commercially is narrow but real. A synthesis error that substitutes one residue for another may be invisible on a low-resolution instrument, present at a few per cent, and chromatographically unresolved from the parent under a fast gradient. The combination of a twelve-minute purity method and a unit-resolution identity check is not a conspiracy; it is simply a pair of tests neither of which is looking in that direction.

Print the spectrum. It is the cheapest available improvement to identity reporting and it consists of reproducing a picture you already have.

The Journal’s request to laboratories, restated annually

Deamidation: 0.98 daltons, and where it hides

Deamidation of asparagine proceeds through a five-membered succinimide intermediate formed by nucleophilic attack of the following residue’s backbone nitrogen on the asparagine side-chain carbonyl. Hydrolysis of the intermediate yields aspartate or isoaspartate, in a ratio typically favouring the isoaspartate form. Glutamine deamidates by an analogous but slower route. The rate depends strongly on pH, temperature, and the identity of the residue immediately following the asparagine, with glycine and serine accelerating it markedly.2

The analytical difficulty is threefold. The mass increase is 0.984 daltons, which requires only modest resolving power to see at low molecular weight and becomes demanding as the peptide gets larger. The aspartate and isoaspartate products are exactly isobaric with one another, so distinguishing them requires either a chromatographic separation that happens to resolve them or a specific enzymatic assay. And deamidated species often elute close to the parent under reversed-phase conditions, so a fast gradient may not separate them either.

The result is a degradation product that is common, that has real consequences for biological activity, that accumulates in storage, and that a certificate produced by a unit-resolution instrument on a twelve-minute gradient is structurally unable to detect. When the Journal describes a certificate as silent on stability, this is a large part of what is meant.

Instrument classes and what each can be asked to support
AnalyserTypical resolving powerTypical mass accuracyCan assign charge from isotope spacing?
Single quadrupole~1,000 (unit)100–500 ppmNo
Linear ion trap2,000–4,00050–200 ppmAt low m/z only
Linear MALDI-TOF500–1,500200–1,000 ppmNo
Reflectron MALDI-TOF10,000–20,0005–50 ppmYes
Quadrupole time-of-flight30,000–60,0001–5 ppmYes
Orbital trap60,000–500,000<1–3 ppmYes
FT-ICR>1,000,000<1 ppmYes
Figures are representative of instruments in general service and are quoted by manufacturers at favourable m/z values; performance at peptide molecular weights is generally lower. Accuracy figures assume routine calibration, and the better end of each range generally requires an internal calibrant.

Oxidation: sixteen daltons, and the one that gets caught

Methionine oxidises to the sulphoxide with a mass increase of 15.995 daltons, and on to the sulphone at a further 15.995. Tryptophan and histidine oxidise by related routes. The chemistry is driven by dissolved oxygen, by trace peroxides in excipients and in some grades of polysorbate, by light, and by transition-metal contamination, and it proceeds in lyophilised material as well as in solution, though more slowly.

Analytically this is the easy case, and it deserves to be described as such in an article otherwise concerned with what cannot be seen. A sixteen-dalton shift is resolvable on essentially any instrument, and the oxidised species is usually chromatographically distinct enough to appear as a separate peak under a reasonably shallow gradient. A spectrum showing a plus-sixteen satellite at a few per cent of the parent intensity is unambiguous evidence of oxidation, and its absence is meaningful evidence of the opposite.

Which is why the Journal’s standing request to laboratories in this market is for the spectrum rather than the verdict. A reproduced spectrum, even at the modest resolution of a routine instrument, allows a reader to look for the plus-sixteen satellite themselves. A conformance statement does not, and cannot be made to. This is the cheapest available improvement to identity reporting in the trade and it consists of printing a picture the laboratory has already produced.

The inversion that weighs nothing

Amino acids other than glycine are chiral, and peptide synthesis is performed with L-configured building blocks. Racemisation during synthesis — most commonly at cysteine, histidine and aspartate residues, and promoted by prolonged base exposure during coupling and deprotection — produces a peptide containing one or more D residues. The resulting molecule has the same elemental composition, the same monoisotopic mass, the same average mass, and the same fragmentation masses as the intended product.

Mass spectrometry cannot detect it. This is not a limitation of any particular instrument; it is a consequence of what the technique measures. Reversed-phase chromatography sometimes separates diastereomeric peptides, and where it does the epimer appears as a shoulder or a satellite peak of unassigned identity — which is one reason a chromatogram with an unexplained minor peak deserves more attention than a purity percentage does. Where the epimer co-elutes, no routine analysis in this market would find it.

Deliberate detection requires chiral amino-acid analysis after total hydrolysis, or digestion with a stereospecific protease that fails to cleave across a D residue, or in some cases ion-mobility separation. None of these is offered as a standard service to private buyers by any of the four testing services this market relies on, and the Journal’s position is that this is a genuine gap rather than a failing on their part: nobody has ever been asked to price it.

The standing rule in this department

This publication applies one rule to every identity claim it reports, and it is worth stating in isolation because it governs the rest. A mass measurement supports a statement about composition. Only a fragmentation or mapping experiment supports a statement about sequence. Where a source says identity was confirmed, we report that a mass was measured, unless we have seen evidence of the second kind.

The rule has consequences we accept. It makes our coverage read as more sceptical than the underlying documents, because the documents claim more than they establish. It occasionally irritates laboratories which have in fact done sequence-level work and have simply not printed it, and the remedy there is a two-line email which we are glad to receive. And it means we cannot describe any research-grade vial in this market as sequence-confirmed, because on the evidence available to us almost none are.

What the rule is not is an accusation. Nothing in this article suggests that vendors are selling material other than what they label, and the Journal has no evidence of that in respect of any company it covers. The claim is narrower and, we think, harder to argue with: the documentation in general circulation does not have the discriminating power that the language on it implies, and the gap between the two is where every avoidable dispute in this market begins.

If this market spent one more pound on identity, where should it go

A fair question, and the Journal’s answer has changed. Our first instinct was to argue for sequence confirmation on every lot, and the arithmetic does not support it: peptide mapping on every batch would raise the analytical cost per vial by a multiple, and the failure mode it protects against — a wholly substituted or permuted sequence — is not the one we see evidence of.

The better allocation, on our present assessment, is orthogonal. Identity by high-resolution intact mass on every lot, at a resolving power sufficient to resolve a one-dalton shift at the parent mass, with the spectrum reproduced. Sequence confirmation once per synthesis campaign rather than once per lot, on the reasoning that the sequence is a property of the process and the lot-to-lot risk is degradation rather than misconstruction. And a chromatographic method shallow enough to separate the deamidated form, because that is the change most likely to have occurred between the certificate and the buyer.

That package is not expensive. Two of the twenty companies in our dossier programme already do something close to the first item, and one has told us it is costed for the second. Whether any of it happens depends on whether buyers ever ask, which is a market question rather than a scientific one and is therefore the harder of the two.

The next piece in this department takes the document rather than the instrument as its subject: what a certificate of analysis contains, what it systematically omits, and how to check one in the time it takes to drink a coffee. Identity is one line on that page, and by the standards of the rest of it, one of the better-behaved ones.

References

  1. “Discrimination of near-isobaric amino acid substitutions in peptides by high-field mass analysis.” Analytical Chemistry. 2016;88(11):5645–5653.
  2. “Asparagine deamidation in synthetic peptides: sequence dependence, kinetics and analytical detection.” Journal of Peptide Science. 2018;24(6):e3092.

Letters to the Editor

3 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

Your article says a matching mass does not confirm a sequence, which is correct, and then rather implies that vendors are trading on the ambiguity. I run analytical services and I would put it differently: we report what we measured, in the words our clients ask for. If the Journal wants the word confirmed retired, write to the buyers, not to us.

W. Stroud, Chattanooga, TN

The Journal replies

That is a fair reallocation of the criticism and we accept it. The word is chosen by whoever commissions the report, and laboratories are answering the question they were paid to answer. Our complaint is with the practice, not with the analysts, and the article should have located it more precisely.

I would add one omission to your six lines: the date and nature of the last calibration. A parts-per-million figure from an instrument last calibrated a fortnight ago is a different claim from one calibrated that morning with an internal standard.

R. Devaney, Ballarat, VIC

The Journal replies

Agreed, and it may be the best suggestion we have received on this subject. It is now a seventh line in the version of the list we send to suppliers, with the note that internal calibration should be stated where it was used.

You state that fourteen of twenty suppliers report an MS identity test. Does that count reports supplied to you on request, or only what appears on the certificate a customer receives?

B. Tejeda, Santo Domingo

The Journal replies

The former, which the table note now says explicitly. The count for what appears on a customer-facing certificate is lower in at least four cases, and we should have separated the two columns rather than merging them.