Vol. 3, No. 6 — June 2026Independent since 2024

TheCompound Journal

Reporting on incretins, compounding & the peptide supply chain

A monthly journal of record.
30 issues · 32 contributors
Not medical advice. We sell nothing.

Method

One method is one opinion

Aggregates dissociate in the mobile phase and are recorded as monomer. Only a size-based separation reports them.

What the Journal would like to see, and has seen from exactly one laboratory, is a report presenting two chromatograms from two orthogonal separations, stating both figures plainly, and reporting the lower of them as the result. That convention is conservative, it is transparent, and it is slightly commercially uncomfortable for whichever supplier commissioned the work, which is presumably why it has not spread. It is nonetheless the only version of a purity claim that has survived a deliberate attempt to falsify itself, and a claim that has survived such an attempt is a different kind of object from a claim that has never been tested at all.

What the instrument does, in order

A pump delivers a mixture of two solvents in a proportion that changes over time under program control. An autosampler injects a measured volume of dissolved sample into that stream. The stream passes through a column packed with particles bearing a bonded hydrophobic ligand, held in a thermostatted compartment. Analytes partition between the mobile phase and the stationary phase; as the organic fraction of the mobile phase rises, each species reaches a composition at which it prefers the mobile phase and leaves the column. A detector at the outlet measures ultraviolet absorbance continuously. Software records the signal and integrates it.

Peptides behave unusually within that framework, in a way worth knowing. Their retention is extremely sensitive to organic composition — much more so than small molecules — which means peptides do not so much elute gradually as leave the column over a narrow composition window. This is why isocratic separation of peptides is impractical and why gradient slope dominates the outcome. It is also why small changes to a gradient programme produce disproportionate changes in resolution.

Every element in the chain is a variable that a method disclosure would specify: column dimensions, particle and pore size, bonded phase, temperature, mobile-phase composition and additive, flow rate, gradient programme, injection volume and sample concentration, detection wavelength and bandwidth. Twelve numbers, all known to the analyst.1

The column: particle, pore and phase

Column choice sets the ceiling on what any gradient can achieve. Three parameters dominate. Particle size governs efficiency: reducing it narrows peaks, and the shift from five-micron to sub-two-micron packings over the past two decades is the reason a modern separation can resolve in ten minutes what once took forty, at the cost of much higher operating pressure and instruments built for it. Superficially porous or core-shell particles achieve much of the same benefit at moderate pressure by shortening the diffusion path.

Pore diameter governs access. The classical hundred-ångström pore was developed for small molecules and becomes restrictive as analyte size rises; for larger peptides a phase with pores in the region of three hundred ångströms allows the molecule to enter the particle and interact with the full bonded surface rather than only the exterior. Using a narrow-pore column for a large peptide produces broad, poorly shaped peaks that are frequently attributed to the sample.

Bonded phase chemistry governs selectivity. Octadecyl silica is the default and covers most peptide work; octyl phases retain less and can help with very hydrophobic sequences; phenyl and polar-embedded phases offer genuinely different selectivity and are therefore candidates for an orthogonal second method. Column dimensions matter too: at constant particle size a longer column gives more resolution and more pressure, and halving the internal diameter quarters the solvent consumption.2

A method reading at 280 nanometres cannot see a fragment that has lost its aromatic residue, however much of it is there.

On detection wavelength

Injection load, linearity and the flattened peak

A detector responds linearly to concentration over a defined range and then stops. Overload the column or saturate the detector with too much sample and the main peak flattens at the top, its apex broadens, and its integrated area no longer represents the quantity present. Since the main peak is the numerator and dominates the denominator, distorting it distorts the purity figure — usually downwards, because the flattened peak loses area relative to a properly loaded one.

There is a competing pressure, and it is the reason overloading happens. Small impurities near the reporting threshold need adequate signal-to-noise to be integrated at all, and the way to raise their signal is to inject more sample. A laboratory hunting for 0.05 per cent impurities is tempted towards a load that compromises the main peak. The correct answer in regulated practice is two injections: a small load for the main peak and a larger one for the related-substances profile, with the results combined.

Column overload is a separate phenomenon from detector saturation and produces a characteristic asymmetric fronting peak. Both are visible on the chromatogram to anybody who is shown it, which is one of several reasons the Journal asks for the trace rather than the number. A purity figure calculated from a distorted main peak is arithmetically correct and analytically meaningless, and the only way to know is to look.3

Detection wavelength and what responds to it
WavelengthPrincipal absorberSees fragments without aromaticsBaseline noiseTypical use
214 nmAmide bondYesHigherPeptide purity and related substances
220 nmAmide bondYesModeratePeptide purity, quieter baseline
254 nmAromatic systemsNoLowSmall-molecule work, legacy detectors
280 nmTrp, Tyr, Phe side chainsNoLowTracking an aromatic residue; not a purity method
Diode array, 200–400 nmAll of the aboveYesMethod-dependentPeak purity assessment, spectral homogeneity
A purity figure generated at 280 nm is not comparable with one generated at 214 nm, and the difference is not a matter of a percentage point. Certificates reading at 280 nm circulate in this market; readers have sent us several.

Reporting, identification and qualification thresholds

Regulated pharmaceutical practice does not treat all small peaks alike. It defines three thresholds. A reporting threshold is the level above which an impurity must be listed in the results. An identification threshold is the level above which its structure must be established. A qualification threshold is the level above which its biological safety must be addressed. The three are set by dose and product class, and the framework converts an argument about small peaks into a documented decision procedure.

The trade has no equivalent. There is no reporting threshold, so an impurity is listed or not according to the software settings; no identification threshold, so nothing is ever identified; and no qualification threshold, because there is no regulatory obligation to qualify anything in a research chemical. The practical consequence is that a certificate reporting 98.6 per cent purity conveys nothing about whether the missing 1.4 per cent is thirty innocuous deletion sequences or one substantial unidentified species.

This is the gap the Journal would most like to see narrowed, and it can be narrowed cheaply. Listing the three largest impurities with their relative retention times and areas costs nothing, requires no additional analysis, and transforms the informational content of the document. One of the twenty companies we track does it. It is not a coincidence that the same company answered every question we put to it about its analytical methods.4

System suitability, the record nobody shows you

Before a laboratory accepts results from a run, it demonstrates that the system was performing. The checks are standard and their names are worth knowing. Resolution between a defined critical pair confirms the separation still separates. Tailing factor confirms peak symmetry within limits, typically not exceeding two. Relative standard deviation of replicate standard injections confirms injection and detection repeatability, commonly required below one or two per cent for an assay. Signal-to-noise at the reporting threshold confirms that small peaks can actually be measured. Theoretical plate count confirms the column has not deteriorated.

A run failing system suitability is discarded, not reported. This is entirely routine, entirely documented, and completely invisible to the recipient of a certificate in this trade. The consequence is that a purity figure arrives with no evidence that the instrument producing it was performing adequately at the time — not because the laboratory has anything to hide, but because the report format never had a field for it.

The Journal asked the four independent services which suitability criteria they apply to a standard peptide purity run. All four apply criteria. Two supplied them in writing. What none of them prints on the report is whether the criteria were met on the specific run that produced your number, which is the only question a reader actually has. A single line reading that system suitability criteria were met would close it.5

1128456280100Purity figure25Wavelength20Chromatogram10Threshold5Gradient0Front windowper cent of companies
Figure. Proportion of the twenty tracked companies disclosing each method parameter on a standard certificate. The purity figure is universal; the four values needed to compare one figure with another are not.

Mass spectrometry, and the limits of a matching mass

Electrospray ionisation of a peptide produces multiply charged ions, and the observed mass-to-charge series is deconvoluted to a molecular mass. Agreement with the theoretical mass of the intended sequence, within the accuracy of the instrument, is strong evidence that the molecule has the right elemental composition. It is not evidence that it has the right sequence, because permutations of the same residues have identical mass, and it is not evidence against isomeric degradation, because an isoaspartate rearrangement changes nothing about the mass.

Fragmentation closes most of that gap. Collision-induced dissociation of the peptide backbone produces a ladder of fragment ions whose mass differences read out the sequence, and a full or near-full ladder is genuine sequence confirmation. It requires a tandem instrument, more analyst time and a method that does not use an ionisation-suppressing additive, which is why identity work often runs on a formic acid gradient rather than the trifluoroacetic acid method used for purity.

The practical reading of a certificate follows. Identity confirmed by mass means the elemental composition matches. Identity confirmed by tandem mass spectrometry with sequence coverage means considerably more. Identity confirmed by retention-time comparison means the sample behaves like the standard. Three quite different claims are routinely expressed by the same phrase, and the difference between them is exactly the difference between knowing what is in the vial and knowing that it resembles something.6

The second method, and what makes it orthogonal

Orthogonality is not a synonym for repetition. Two runs of the same method differ only in random variation. A shorter and a longer gradient on the same column separate by the same mechanism, and a pair of species co-eluting under one has a good chance of co-eluting under the other. Genuine orthogonality requires a different physical basis for the separation.

For peptides the practical options are well established. Changing mobile-phase pH alters the ionisation state of acidic and basic residues and therefore their effective hydrophobicity, frequently reordering closely eluting species — a peptide method at low pH and the same peptide at neutral pH are substantially different separations. Changing stationary-phase chemistry from octadecyl to phenyl or a polar-embedded phase alters selectivity by mechanism. Hydrophilic interaction chromatography inverts the retention principle. Ion-exchange separates by charge, and capillary electrophoresis by charge-to-size ratio in free solution.

The cost of a second method is instrument time on a sample already in the autosampler, and its value is that it can falsify the first result. Where the two agree, confidence rises substantially. Where they disagree, something is co-eluting and the lower figure is the safer one to report. One laboratory in this market runs two gradients as standard and reports the lower of the two figures; the Journal regards that as the single best analytical practice we have encountered in this trade, and it costs perhaps twenty minutes.7

A method that survives ruggedness testing is one that works because of its design choices, not because of luck.

Analytical method validation practice

Five things a purity figure cannot tell you

First, how much peptide is in the vial. Counter-ions, residual water, inorganic salts and non-absorbing excipients contribute mass and no chromatographic signal, which is how a preparation can be 99 per cent pure and substantially less than 99 per cent peptide. Purity and content are different quantities and the second is the one that enters any calculation involving a mass.

Second, whether anything is aggregated. Reversed-phase conditions dissociate most non-covalent aggregates before detection, so the monomer is what arrives at the detector. Only a size-based separation reports high molecular weight species.

Third, whether the sequence is correct. Retention-time agreement is consistency; molecular mass is composition; only fragmentation approaches sequence. Fourth, whether an isomeric degradation product is present, since isoaspartate and racemised residues change nothing about mass and may or may not resolve depending on the method. Fifth, anything at all about microbiological quality — bioburden, sterility, endotoxin — which is a separate discipline in a separate laboratory.

Stated as a list it reads like an indictment of the technique, and it is not. Reversed-phase chromatography answers its own question superbly. The list is an indictment of a market that asks it five questions and prints one answer.

Four laboratories, one lot: what was disclosed alongside the number
LaboratoryPurityGradient disclosedWavelengthThresholdChromatogram supplied
W99.1%Run time only220 nmNot statedYes
X98.5%Full programme214 nm0.10%Yes
Y97.6%Full programme214 nm0.05%Yes, two
Z98.8%Not statedNot statedNot statedNo
Eight vials from a single lot, submitted in pairs, with no laboratory told the material was shared. Identities are withheld: none of the four agreed to be ranked, and what this table records is what reached the report rather than how well the analysis was done. Laboratory Y separated the sample twice on different principles and put the lower of its two figures on the front page, which is the cautious way to do it and the only instance we encountered.

Four values that would make purity figures comparable

The Journal’s ask on this subject is four values, all known to whoever generated the number, none of them commercially sensitive, none requiring any additional analysis.

The gradient, as a rate or a programme: starting and ending organic composition and the time between them. The detection wavelength. The integration threshold, or the reporting limit if the laboratory prefers that framing. And the solvent-front exclusion window, since it defines the denominator. Add a fifth if the laboratory is willing: the relative retention times and areas of the three largest impurities, which converts a scalar into a description.

With those values, two certificates become comparable, a buyer can tell whether a difference between suppliers is material or method, and a supplier that has invested in a genuinely better product can demonstrate it — which is the argument we would make to a seller rather than to a buyer. Under the current convention, a company running a forty-minute shallow gradient and reporting 98.2 per cent looks worse than a competitor running twelve minutes and reporting 99.4, and there is no mechanism by which the first can show a buyer why. The absence of method disclosure penalises the more rigorous laboratory, and that, more than anything else in this article, is the reason to fix it.

A note on method and sourcing

The compendial and regulatory material in this piece is taken from the current general chapters on chromatography and on validation of compendial procedures, from the European Pharmacopoeia chapters on liquid chromatography and on chromatographic separation techniques, and from the harmonised guidelines on analytical validation, on impurities and on specifications for biotechnological products, all read in the original. The separation science is drawn from the chromatography literature, with the peptide-specific behaviour cited where it differs from small-molecule practice.

Where the Journal reports a number it obtained, it states the number of vials, the number of laboratories, whether the vials came from one lot, whether the laboratories knew, and what method parameters were disclosed to us. Where we quote a figure from a certificate we state whether the method was disclosed on it. Where a laboratory or a company answered our questions we distinguish an answer from a refusal and a refusal from a non-response.

Nothing in this department is a recommendation to buy, use or avoid anything. The compounds referred to are sold for research use only and are not approved for human use in any jurisdiction. Corrections and disputes go to standards@compoundjournal.com. Readers with certificates or chromatograms they would like read should write to letters@compoundjournal.com; we do not identify the source of anything sent to us, and we do not publish a reader’s name without permission.

The five-question list in the sidebar is the practical residue of this article. Gradient, wavelength, threshold, standard, second method. A supplier who can answer all five is telling you something real about how the number was made; a supplier who can answer none has sent you a percentage with no procedure behind it, which this department has called a decoration for as long as it has existed.

References

  1. European Pharmacopoeia. Chapter 2.2.29 — Liquid Chromatography. Council of Europe, Strasbourg.
  2. “Stationary phase and pore size selection for reversed-phase separation of peptides and small proteins.” Journal of Chromatography A. 2017;1523:2–18.
  3. United States Pharmacopeia. General Chapter ⟨621⟩ Chromatography. USP–NF, Rockville, MD.
  4. International Council for Harmonisation. Q3A(R2): Impurities in New Drug Substances. 2006.
  5. International Council for Harmonisation. Q2(R2): Validation of Analytical Procedures. 2023.
  6. “Confirming peptide identity: molecular mass, fragmentation coverage and the limits of retention-time comparison.” Journal of Peptide Science. 2019;25(8):e3195.
  7. “Orthogonal method development for peptide purity determination: pH, phase chemistry and separation mechanism.” Journal of Chromatography A. 2020;1618:460873.

Letters to the Editor

1 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

Your worked example varies gradient and threshold together and reports a 1.8-point spread. Which of the two contributed more? The article does not say, and the answer matters for what you are asking suppliers to disclose first.

R. Cadogan, Bridgetown

The Journal replies

Gradient, by roughly two to one in our four conditions: holding the threshold at 0.10 per cent, lengthening the gradient cost 0.9 points, while holding the gradient and tightening the threshold cost 0.4 to 0.9 depending on which gradient. We should have printed that decomposition in the table and it now appears in the note. If a supplier will disclose only one value, it should be the gradient.

Related coverage

Laboratory Notebook

Gauge, pain and flow rate

Gauge affects pain and flow rate rather than depth. A finer needle is more comfortable and slower, and with a viscous solution the difference is noticeable.

Perpetua Nwachukwu·1 Aug 2024·8 min