The impurity hiding underneath the parent peak

The most useful single addition to any purity determination is a second separation under a different pH or on a different stationary phase, and the second-most useful is a mass spectrum.

The reason it matters is that co-elution is the failure mode a single method cannot detect. A chromatogram showing one sharp, symmetrical peak and a purity of 99.3 per cent is consistent with a clean preparation and equally consistent with a preparation in which a related species elutes under the parent. Peak-purity assessment from a diode-array detector helps and is not conclusive. A second separation on a different principle is the practical answer, and it is cheap: the sample is already in the vial, the second run costs instrument time.

Reporting, identification and qualification thresholds

Regulated pharmaceutical practice does not treat all small peaks alike. It defines three thresholds. A reporting threshold is the level above which an impurity must be listed in the results. An identification threshold is the level above which its structure must be established. A qualification threshold is the level above which its biological safety must be addressed. The three are set by dose and product class, and the framework converts an argument about small peaks into a documented decision procedure.

The trade has no equivalent. There is no reporting threshold, so an impurity is listed or not according to the software settings; no identification threshold, so nothing is ever identified; and no qualification threshold, because there is no regulatory obligation to qualify anything in a research chemical. The practical consequence is that a certificate reporting 98.6 per cent purity conveys nothing about whether the missing 1.4 per cent is thirty innocuous deletion sequences or one substantial unidentified species.

This is the gap the Journal would most like to see narrowed, and it can be narrowed cheaply. Listing the three largest impurities with their relative retention times and areas costs nothing, requires no additional analysis, and transforms the informational content of the document. One of the twenty companies we track does it. It is not a coincidence that the same company answered every question we put to it about its analytical methods.1

Mass spectrometry, and the limits of a matching mass

Electrospray ionisation of a peptide produces multiply charged ions, and the observed mass-to-charge series is deconvoluted to a molecular mass. Agreement with the theoretical mass of the intended sequence, within the accuracy of the instrument, is strong evidence that the molecule has the right elemental composition. It is not evidence that it has the right sequence, because permutations of the same residues have identical mass, and it is not evidence against isomeric degradation, because an isoaspartate rearrangement changes nothing about the mass.

Fragmentation closes most of that gap. Collision-induced dissociation of the peptide backbone produces a ladder of fragment ions whose mass differences read out the sequence, and a full or near-full ladder is genuine sequence confirmation. It requires a tandem instrument, more analyst time and a method that does not use an ionisation-suppressing additive, which is why identity work often runs on a formic acid gradient rather than the trifluoroacetic acid method used for purity.

The practical reading of a certificate follows. Identity confirmed by mass means the elemental composition matches. Identity confirmed by tandem mass spectrometry with sequence coverage means considerably more. Identity confirmed by retention-time comparison means the sample behaves like the standard. Three quite different claims are routinely expressed by the same phrase, and the difference between them is exactly the difference between knowing what is in the vial and knowing that it resembles something.2

Retention-time agreement is consistency. Molecular mass is composition. Only fragmentation approaches sequence.

On three claims that share one phrase

The second method, and what makes it orthogonal

Orthogonality is not a synonym for repetition. Two runs of the same method differ only in random variation. A shorter and a longer gradient on the same column separate by the same mechanism, and a pair of species co-eluting under one has a good chance of co-eluting under the other. Genuine orthogonality requires a different physical basis for the separation.

For peptides the practical options are well established. Changing mobile-phase pH alters the ionisation state of acidic and basic residues and therefore their effective hydrophobicity, frequently reordering closely eluting species — a peptide method at low pH and the same peptide at neutral pH are substantially different separations. Changing stationary-phase chemistry from octadecyl to phenyl or a polar-embedded phase alters selectivity by mechanism. Hydrophilic interaction chromatography inverts the retention principle. Ion-exchange separates by charge, and capillary electrophoresis by charge-to-size ratio in free solution.

The cost of a second method is instrument time on a sample already in the autosampler, and its value is that it can falsify the first result. Where the two agree, confidence rises substantially. Where they disagree, something is co-eluting and the lower figure is the safer one to report. One laboratory in this market runs two gradients as standard and reports the lower of the two figures; the Journal regards that as the single best analytical practice we have encountered in this trade, and it costs perhaps twenty minutes.3

Method disclosure across twenty companies’ standard certificates
Disclosed itemOn standard certificateOn requestNot available
Purity figure2000
Method named as HPLC2000
Detection wavelength569
Gradient programme or rate1514
Integration threshold2315
Solvent-front exclusion window0218
Three largest impurities listed1118
Chromatogram attached479
Compiled from standard release documentation and from a written questionnaire sent twice, four weeks apart. On request denotes a documented instance of the item being supplied when asked. Where a supplier attaches an independent laboratory report rather than transcribing a figure, the disclosure is credited to the certificate.

Five things a purity figure cannot tell you

First, how much peptide is in the vial. Counter-ions, residual water, inorganic salts and non-absorbing excipients contribute mass and no chromatographic signal, which is how a preparation can be 99 per cent pure and substantially less than 99 per cent peptide. Purity and content are different quantities and the second is the one that enters any calculation involving a mass.

Second, whether anything is aggregated. Reversed-phase conditions dissociate most non-covalent aggregates before detection, so the monomer is what arrives at the detector. Only a size-based separation reports high molecular weight species.

Third, whether the sequence is correct. Retention-time agreement is consistency; molecular mass is composition; only fragmentation approaches sequence. Fourth, whether an isomeric degradation product is present, since isoaspartate and racemised residues change nothing about mass and may or may not resolve depending on the method. Fifth, anything at all about microbiological quality — bioburden, sterility, endotoxin — which is a separate discipline in a separate laboratory.

Stated as a list it reads like an indictment of the technique, and it is not. Reversed-phase chromatography answers its own question superbly. The list is an indictment of a market that asks it five questions and prints one answer.

A note on method and sourcing

The compendial and regulatory material in this piece is taken from the current general chapters on chromatography and on validation of compendial procedures, from the European Pharmacopoeia chapters on liquid chromatography and on chromatographic separation techniques, and from the harmonised guidelines on analytical validation, on impurities and on specifications for biotechnological products, all read in the original. The separation science is drawn from the chromatography literature, with the peptide-specific behaviour cited where it differs from small-molecule practice.

Where the Journal reports a number it obtained, it states the number of vials, the number of laboratories, whether the vials came from one lot, whether the laboratories knew, and what method parameters were disclosed to us. Where we quote a figure from a certificate we state whether the method was disclosed on it. Where a laboratory or a company answered our questions we distinguish an answer from a refusal and a refusal from a non-response.

Nothing in this department is a recommendation to buy, use or avoid anything. The compounds referred to are sold for research use only and are not approved for human use in any jurisdiction. Corrections and disputes go to standards@compoundjournal.com. Readers with certificates or chromatograms they would like read should write to letters@compoundjournal.com; we do not identify the source of anything sent to us, and we do not publish a reader’s name without permission.

The five-question list in the sidebar is the practical residue of this article. Gradient, wavelength, threshold, standard, second method. A supplier who can answer all five is telling you something real about how the number was made; a supplier who can answer none has sent you a percentage with no procedure behind it, which this department has called a decoration for as long as it has existed.

References

  1. International Council for Harmonisation. Q3A(R2): Impurities in New Drug Substances. 2006.
  2. “Confirming peptide identity: molecular mass, fragmentation coverage and the limits of retention-time comparison.” Journal of Peptide Science. 2019;25(8):e3195.
  3. “Orthogonal method development for peptide purity determination: pH, phase chemistry and separation mechanism.” Journal of Chromatography A. 2020;1618:460873.

Letters to the Editor

4 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

Your article argues for two orthogonal methods and reports the lower figure, but the people running a single twelve-minute method have a cost story you do not address. A full orthogonal pair doubles the turnaround and at least doubles the cost, which is why the market does not do it. The criticism of method disclosure is fair. The criticism that a single method is wrong is unfair to the constraints people operate under.

P. Kovalenko, Lviv

The Journal replies

We are careful to say that a second method costs instrument time on a sample already in the autosampler, which is substantially less than twice the turnaround, but you are right that we underweight the commercial reality that a buyer setting a budget for testing is trading thoroughness for speed and price. Where we would push back is that those constraints are not technical or regulatory ones. They are market ones, and markets can change if enough buyers demand it.

The table showing what each method can detect is valuable but incomplete. You show no row for C-terminal truncation or N-terminal truncation as distinct phenomena. These are not rare, and they often elute differently depending on which end is missing. A generic gradient might resolve them; an improperly designed orthogonal method might not. The capability matrix should separate these cases.

D. Ramkissoon, Port of Spain

On the section about diode-array detection and peak purity, I would add that true peak purity assessment requires library matching or at least spectral comparison across the peak width. A homogeneous spectrum tells you the peak is probably pure. A spectrum that shifts across the peak tells you it is not, and that information closes a gap the article identifies correctly.

R. Anand, Pune

You write that only one laboratory attached its chromatogram to the private buyer report. That was probably us. We started doing it five years ago because the PDF seemed incomplete without it. It costs us nothing to add — the instrument generates it automatically — and it solves exactly the dispute-resolution problem you describe. More laboratories should do it, and the reason they do not is not technical.

C. Adeoti, Ibadan

The Journal replies

That is generous of you to say. The technical barrier is near zero, and if enough laboratories began printing them, it would force the convention to change across the market. It is an example of something that costs one actor almost nothing but creates value for everyone, and it is precisely the kind of thing that can shift a trade practice when a few leaders move first.