Xianning export categories for peptide intermediates were quietly renamed
Follow the resin, not the catalogue.
TheCompound Journal
Reporting on incretins, compounding & the peptide supply chain
Mass spectrometry
The ionisation method determines the charge states you see, the adducts you must account for, and the modifications you might destroy in the process.
Ask a laboratory which ionisation source produced a spectrum and you learn a great deal about the spectrum before you have looked at it. Electrospray on a modern instrument will give a peptide of four thousand daltons a family of multiply charged ions and a mass accuracy in the low parts per million. A linear MALDI time-of-flight will give a single predominantly singly charged ion and an accuracy measured in hundreds of parts per million. Both can answer the question, is this the right compound. Only one of them can answer the question, is this the right compound and nothing very close to it.
It is worth being exact about what a mass spectrometer does, because the imprecision propagates. The instrument generates ions from a sample, separates them according to the ratio of their mass to their charge, and counts them at a detector. The horizontal axis of every spectrum is mass-to-charge, conventionally written m/z and expressed in thomsons or in dimensionless units depending on the vendor’s software. Nothing is weighed. Nothing is measured against a reference mass in the sense that a balance measures against a calibration weight.
What follows from this is that every molecular weight on every certificate of analysis in this market is a calculated quantity, derived from a measured m/z by assigning a charge and subtracting the mass contribution of whatever adducted to the molecule to give it that charge — usually protons, sometimes sodium, occasionally potassium or ammonium. The assignment is normally straightforward and normally correct. It is nonetheless an assignment, and when it goes wrong it goes wrong by an integer factor, which is the kind of error that produces confident nonsense rather than a plausible discrepancy.
The practical consequence for a reader is a habit: when a mass figure appears, ask what was observed and what was inferred. A report that gives both — the m/z, the charge, and the derived neutral mass — has answered the question before it was asked.
Electrospray ionisation works by pumping a solution of the analyte through a fine capillary held at a potential of a few kilovolts relative to the instrument’s entrance. The liquid emerging from the tip forms a cone and then a jet of charged droplets. As solvent evaporates the droplets shrink, the charge density on their surfaces rises, and at the point where electrostatic repulsion exceeds surface tension they fission into smaller droplets. Repeat this enough times and what is left is a bare, charged analyte ion in the gas phase.
Because the charge is acquired in solution and retained through desolvation, a peptide with several basic residues will carry several protons, and the population of ions reaching the analyser is distributed across charge states. This is the defining characteristic of electrospray spectra and the reason they look bewildering to a first-time reader: a single pure compound produces four or five prominent peaks, none of them at the molecular weight.
The distribution is not noise. It carries information about the number of accessible basic sites and about the conformational state of the molecule in solution, and it shifts predictably with mobile-phase composition and pH. It also has a practical advantage that matters for identity work: dividing the mass by three or four brings a large peptide into the range where instruments achieve their best resolving power and accuracy.1
A matching mass has established that the vial contains something of the same elemental composition. That is not identity, and it is not close.
The standing rule in this departmentMatrix-assisted laser desorption ionisation takes a different route. The peptide is mixed with a large molar excess of a small organic compound that absorbs strongly at the laser wavelength — α-cyano-4-hydroxycinnamic acid and sinapinic acid are the usual choices for peptides and proteins respectively — and the mixture is dried on a metal target. A pulsed ultraviolet laser strikes the crystal, the matrix absorbs the energy, and a plume of matrix and analyte is ejected into the vacuum with the analyte largely intact and mostly singly protonated.
Two consequences follow. First, MALDI spectra are simple: one predominant ion per compound, at the molecular weight plus one proton, which makes them easy to read and easy to print on a certificate. Second, MALDI is markedly more tolerant of salts, buffers and heterogeneous samples than electrospray, which is why it survives in routine synthesis monitoring where electrospray would require a chromatographic clean-up first.
The trade-offs are equally real. The matrix produces intense chemical background below roughly a thousand daltons, which obscures small fragments. Ion yield varies between compounds and between spots on the same target, making MALDI a poor quantitative technique. And the achievable mass accuracy on a linear instrument at peptide molecular weights is measured in hundreds of parts per million unless a reflectron and delayed extraction are in use.2
| Analyser | Typical resolving power | Typical mass accuracy | Can assign charge from isotope spacing? |
|---|---|---|---|
| Single quadrupole | ~1,000 (unit) | 100–500 ppm | No |
| Linear ion trap | 2,000–4,000 | 50–200 ppm | At low m/z only |
| Linear MALDI-TOF | 500–1,500 | 200–1,000 ppm | No |
| Reflectron MALDI-TOF | 10,000–20,000 | 5–50 ppm | Yes |
| Quadrupole time-of-flight | 30,000–60,000 | 1–5 ppm | Yes |
| Orbital trap | 60,000–500,000 | <1–3 ppm | Yes |
| FT-ICR | >1,000,000 | <1 ppm | Yes |
| Figures are representative of instruments in general service and are quoted by manufacturers at favourable m/z values; performance at peptide molecular weights is generally lower. Accuracy figures assume routine calibration, and the better end of each range generally requires an internal calibrant. | |||
The choice between the two techniques is not a matter of quality but of question. A synthesis chemist watching a coupling proceed wants a fast, salt-tolerant check that the chain has grown by the expected residue, and MALDI on a bench instrument answers that in minutes. An analytical laboratory asked whether a submitted vial contains the labelled compound and nothing closely related to it needs the resolving power and the accuracy that electrospray into a high-field analyser provides, coupled to a chromatographic separation so that species which co-elute can at least be assigned to retention times.
Both appear in this market, and reports rarely distinguish them. That matters because the two techniques have different blind spots. MALDI can induce loss of labile modifications during desorption, so a phosphorylated or otherwise fragile species may be under-represented. Electrospray suppresses ionisation of some analytes in the presence of others, so a minor component of a mixture may be absent from a spectrum in which it is genuinely present.
A certificate stating the source therefore tells a reader which class of error to consider. The Journal has stopped asking suppliers for more testing and started asking them for this line instead, on the grounds that it costs nothing and changes what the existing test can be said to support.
For a peptide of neutral monoisotopic mass M observed as a protonated ion carrying z protons, the mass-to-charge ratio is (M + z × 1.00728) divided by z, where 1.00728 is the mass of a proton — the mass of a hydrogen atom less the mass of an electron, a distinction that matters at parts-per-million accuracy and not at all below it.
Run this for a peptide of average mass 4113.58. The singly protonated ion appears at 4114.59. The doubly protonated ion appears at 2057.80, the triply at 1372.20, the quadruply at 1029.40 and the quintuply at 823.72. All five describe the same molecule. A reader shown only the fourth of those figures, without a charge assignment, would reasonably conclude the vial contained a peptide of about a thousand daltons.
Inverting the calculation is how the neutral mass is recovered: multiply the observed m/z by the charge and subtract z proton masses. Doing this for two or three charge states from the same spectrum and finding agreement to within the instrument’s stated accuracy is the standard internal consistency check, and it is the check that catches a misassigned charge. A single m/z with a single assumed charge has no such redundancy, which is one reason electrospray with a visible charge-state envelope is more informative than a single MALDI peak even when both instruments are equally well calibrated.
Mass accuracy is the difference between the measured mass and the true mass, expressed in parts per million of the measured value. It depends on calibration, on the stability of the instrument’s electronics and temperature, on the number of ions arriving at the detector, and on whether an internal calibrant was co-analysed with the sample. It is not a fixed property of an instrument; it is a property of a measurement made on an instrument on a particular day.
Certificates in this market seldom state a tolerance at all. Where they do, the figure is usually expressed in daltons rather than parts per million and is generous: ±0.5 or ±1.0 dalton is common, which at incretin molecular weights corresponds to 120 to 240 parts per million and is achievable on almost any instrument sold in the last thirty years. A tolerance that no plausible measurement could fail is not an acceptance criterion. It is a formality.
What a meaningful criterion looks like is not mysterious. State the theoretical mass and its convention, state the observed mass, state the deviation in parts per million, and state the limit above which the result would have been reported as non-conforming. Four numbers, all of them already known to the analyst. The compendial framework for validating an analytical procedure asks for exactly this kind of specificity about what a test can discriminate, and the framework predates this market by decades.3
Every mass spectrometer is calibrated against a mixture of compounds of known exact mass, and every mass spectrometer drifts away from that calibration afterwards. The rate depends on the analyser type, on ambient temperature stability, and in trapping instruments on the number of ions in the trap: space-charge effects shift apparent masses in a manner that depends on how much sample was injected.
External calibration means the calibrant was run separately, before or after the samples. It is simple, it is what most routine work uses, and it is vulnerable to everything that happens between the calibration and the sample. Internal calibration means a compound of known mass was present in the same spectrum as the analyte, so the correction is applied to the measurement rather than to the instrument. Internal calibration is the reason sub-part-per-million figures are achievable at all, and it is the difference between a stated accuracy and a demonstrated one.
None of this is exotic or contested; it is ordinary laboratory practice, described in accreditation requirements as part of metrological traceability and in the pharmacopoeial chapters as part of system suitability.4 The reason it belongs in an article aimed at buyers is that it explains why two competent laboratories analysing the same vial on the same class of instrument can differ by tens of parts per million, and why the honest response to such a difference is to ask about calibration rather than about honesty.
A mass spectrometer does not weigh anything. It measures the trajectory of an ion, and everything else on the certificate is an inference.
Callum Brathwaite, Analytical Chemistry CorrespondentPeptide mapping is the standard method by which the primary structure of a peptide or protein product is verified. The material is digested with a protease of defined specificity — trypsin cleaving after lysine and arginine, Lys-C after lysine alone, chymotrypsin after aromatic residues — and the resulting fragments are separated by reversed-phase chromatography with mass detection. Each fragment’s observed mass is matched against the masses predicted from the expected sequence, and the fraction of the chain accounted for is reported as sequence coverage.
A digest that returns every predicted fragment at the predicted mass is a far stronger identity statement than an intact mass, because it constrains the order of residues in a way that an intact measurement does not: a permuted sequence generally produces different cleavage products. It is not complete on its own, because a fragment mass is subject to the same permutation ambiguity in miniature, which is why serious mapping proceeds to a second stage of mass analysis on the fragments themselves.
For synthetic peptides of thirty to forty residues, mapping is straightforward chemistry and unremarkable chromatography, and the reason it does not appear on certificates in this market is cost and turnaround rather than difficulty. The regulatory expectation for a peptide product characterised as a biotechnological article treats structural confirmation of this kind as a matter of routine rather than of specialism.56
| Change | Mass shift (Da) | Shift (ppm at 4000 Da) | Resolving power required |
|---|---|---|---|
| Oxidation (one O added) | +15.995 | 3999 | ~250 |
| Deamidation of Asn or Gln | +0.984 | 246 | ~4100 |
| Disulphide formation | −2.016 | 504 | ~2000 |
| Pyroglutamate formation | −18.011 | 4503 | ~220 |
| TFA adduct | +113.993 | 28498 | ~35 |
| Gln replaced by Lys | −0.036 | 9 | ~110000 |
| Leu replaced by Ile | 0.000 | 0 | not resolvable |
| L to D inversion | 0.000 | 0 | not resolvable |
| Required resolving power estimated as the parent mass divided by the mass shift, which is the minimum needed to present the two species as separate peaks; in practice a factor of two above this figure is needed for reliable quantitation of the minor species. | |||
Methionine oxidises to the sulphoxide with a mass increase of 15.995 daltons, and on to the sulphone at a further 15.995. Tryptophan and histidine oxidise by related routes. The chemistry is driven by dissolved oxygen, by trace peroxides in excipients and in some grades of polysorbate, by light, and by transition-metal contamination, and it proceeds in lyophilised material as well as in solution, though more slowly.
Analytically this is the easy case, and it deserves to be described as such in an article otherwise concerned with what cannot be seen. A sixteen-dalton shift is resolvable on essentially any instrument, and the oxidised species is usually chromatographically distinct enough to appear as a separate peak under a reasonably shallow gradient. A spectrum showing a plus-sixteen satellite at a few per cent of the parent intensity is unambiguous evidence of oxidation, and its absence is meaningful evidence of the opposite.
Which is why the Journal’s standing request to laboratories in this market is for the spectrum rather than the verdict. A reproduced spectrum, even at the modest resolution of a routine instrument, allows a reader to look for the plus-sixteen satellite themselves. A conformance statement does not, and cannot be made to. This is the cheapest available improvement to identity reporting in the trade and it consists of printing a picture the laboratory has already produced.
The Journal has settled on a short list, arrived at by writing to laboratories and asking what they could supply without additional work. Six lines. The ionisation source and mode. The analyser, named by class at minimum and by model preferably. The theoretical mass, with the convention stated as monoisotopic or average. The observed mass, with the charge state from which it was derived. The deviation, expressed in parts per million. And the acceptance criterion that was applied.
Every one of those is in front of the analyst at the moment the report is generated. None is commercially sensitive. Together they convert a verdict into a measurement, because they allow a reader to determine what the test could have detected and what it could not. A document carrying those six lines can be assessed by somebody who has never seen the sample; a document reading “MS: conforms” cannot be assessed at all, by anybody, including the person who wrote it.
The compendial approach to identity testing is built on the same three elements — a technique, a reference, and a criterion — and asks for them to be stated because a test whose discriminating power is undocumented has not been validated in any meaningful sense.7 We are not asking this market to become a regulated one. We are asking it to print what it already knows.
This publication applies one rule to every identity claim it reports, and it is worth stating in isolation because it governs the rest. A mass measurement supports a statement about composition. Only a fragmentation or mapping experiment supports a statement about sequence. Where a source says identity was confirmed, we report that a mass was measured, unless we have seen evidence of the second kind.
The rule has consequences we accept. It makes our coverage read as more sceptical than the underlying documents, because the documents claim more than they establish. It occasionally irritates laboratories which have in fact done sequence-level work and have simply not printed it, and the remedy there is a two-line email which we are glad to receive. And it means we cannot describe any research-grade vial in this market as sequence-confirmed, because on the evidence available to us almost none are.
What the rule is not is an accusation. Nothing in this article suggests that vendors are selling material other than what they label, and the Journal has no evidence of that in respect of any company it covers. The claim is narrower and, we think, harder to argue with: the documentation in general circulation does not have the discriminating power that the language on it implies, and the gap between the two is where every avoidable dispute in this market begins.
Readers who take one thing from this piece should take the arithmetic. Isotope spacing is one over the charge. Deamidation is one dalton, oxidation is sixteen, and stereochemical inversion is nothing at all. A tolerance of ±1 dalton on a four-thousand-dalton peptide is two hundred and forty parts per million and excludes almost nothing worth excluding. Those four facts are sufficient to read most of the identity claims in circulation, and they fit on the back of an envelope.
Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.
Sixteen years in a peptide plant and I have never once been asked by a customer which ionisation source we used. I have been asked hundreds of times for a purity figure to one more decimal place.
— N. Ó Broin, Sligo
The claim that a reproduced spectrum is worth more than any number in the document seems overstated. Most buyers cannot read a spectrum, and a printed image invites false confidence rather than scrutiny.
— F. Duquesne, Lyon
Partly conceded. A spectrum is worth more to a reader who can read one, and this department exists partly to increase that number. But it is also an artefact that can be checked by a third party later, which a bare verdict is not, and that alone justifies printing it.
On your point about D-amino acids: chiral amino-acid analysis after hydrolysis is not exotic and several contract laboratories offer it. The obstacle is that hydrolysis itself racemises a few per cent of most residues, so the method has a blank problem, and interpreting a low-level D content is genuinely difficult rather than merely expensive.
— S. Nortje, Stellenbosch
An important qualification and we are glad to have it. The article implied the barrier was commercial when a substantial part of it is methodological. Recorded, and the section has been rewritten accordingly.
Follow the resin, not the catalogue.
Two signatures — one who performed the work, one who approved its release — are the ordinary regulated convention and are almost unknown here.
The arithmetic of significant figures, applied to a document that routinely reports four of them.
The two ionisation techniques in general use produce different charge distributions, different adducts and different failure modes. Certificates almost never say which was…
A 4 mm needle at ninety degrees without a skin pinch is adequate for essentially all adults. The persistence of 12.7 mm needles in this market is habit, not reasoning.
What the renal and hepatic outcome programmes actually measured, and over what duration.