Independent analysis puts a SWB dulaglutide lot at 98.6%, against 96.9% on the certificate
The result is unremarkable. What the report omits is not.
TheCompound Journal
Reporting on incretins, compounding & the peptide supply chain
Incentives
Two years ago we ran an anonymised version of this comparison and promised a named one. This is it, with every method printed in full.
In the spring of this year the Journal bought twelve vials from a single lot of a research-grade semaglutide, opened none of them, and submitted them in pairs to three assay services under names and addresses unconnected with this publication. Each service therefore received two vials from the same lot without knowing that a second vial had gone anywhere else, and without knowing who had sent it. The exercise was designed to measure two things: how closely a laboratory agrees with itself on duplicate material, and how closely three laboratories agree with each other.
The clearest case of the conflict is also the most common. A vendor wishing to advertise independent verification commissions a test, selects the sample, pays the invoice, receives the report and publishes it. Each of those steps is ordinary commerce and none of them is improper. Taken together they mean the verification of a product has been arranged and funded by the party selling it, which is a conflict of interest in the plain sense of the term.
It is worth being exact about what the conflict does and does not imply. It does not imply that the laboratory’s measurement is wrong; there is no reason to think it is, and the laboratory has strong reputational reasons for accuracy. It does not imply that the vendor is dishonest. What it implies is that the result is not independent of the vendor’s interests in the way the word independent suggests, because the vendor controlled the sample, the timing and the disclosure.
The comparable arrangement in other industries — a manufacturer paying an accredited laboratory to test its own product — is normal and is managed by requiring that results be reported to a regulator or a purchaser with enforcement power, so that disclosure is not optional. Neither exists here. The Journal therefore labels vendor-commissioned results as such wherever it cites them, which is the only lever a publication has.
Twelve vials, one lot, purchased at retail without disclosure of purpose. Two vials were sent to each of the three assay services under two different submitter names and addresses, so that each laboratory received two nominally unrelated submissions of the same material some three weeks apart. Six further vials were retained. Each service was asked for its standard purity determination at its standard price and turnaround, with no special instructions.
The design tests two distinct quantities that the trade conflates. Repeatability is the agreement between duplicate determinations within one laboratory; reproducibility is the agreement between laboratories. Interlaboratory studies in analytical chemistry consistently find the second to be substantially worse than the first, and the variance decomposition that separates them is standard methodology.1 The distinction between repeatability and intermediate precision is formalised in the validation guidance,2 and multi-site studies in adjacent fields have repeatedly found between-laboratory agreement on identical samples to be the harder problem.3
All three services were informed after the fact, before publication, and each was given the opportunity to comment on its own method as printed and on the comparison as a whole. All three responded. Two supplied additional method detail that has been incorporated. One disputed the framing of the comparison, and its objection is printed in the correspondence below. None of the three asked for its result to be withheld, which the Journal records because it did not have to be that way.
A badge is not evidence. It is an assertion that evidence exists, offered without the particulars that would let anybody assess it.
The Journal’s standing positionWithin-laboratory repeatability was good. The two determinations from each service agreed to within 0.3 percentage points in every case, and to within 0.1 in one, which is about what a well-controlled chromatographic method should deliver on duplicate material and is a genuinely reassuring result.
Between-laboratory reproducibility was another matter. The three services returned figures spanning 2.1 percentage points on material from one lot. Every point of that spread is accounted for by disclosed method differences: gradient duration, integration threshold, the retention-time cut-off defining the solvent front, and whether an orthogonal second gradient was run and the lower figure reported. Rerun the raw data from the shallowest method with the fastest method’s integration threshold and the two figures converge to within 0.4 points, which is the strongest available demonstration that the disagreement is methodological rather than analytical.
Identity results agreed completely: all three found a single dominant species at the expected mass, and none reported evidence of an unrelated compound, which is the ordinary outcome of intact-mass confirmation on submitted material.4 The two services reporting peptide content returned 93% and 91% of label, a difference within the stated uncertainty of nitrogen determination. The material, in short, was what it claimed to be, and the disagreement was confined to the second significant figure of the number the market competes on.5
| Service | Gradient | Detection | Integration threshold | Result A | Result B |
|---|---|---|---|---|---|
| Service 1 | 15 min generic peptide | 220 nm | not stated | 98.9% | 99.0% |
| Service 2 | 30 min | 214 nm | 0.10% | 98.1% | 98.4% |
| Service 3 | 40 min shallow + orthogonal | 214 nm | 0.05% | 96.8% | 96.9% |
| Twelve vials from one retail lot of a research-grade semaglutide, submitted in pairs three weeks apart under unconnected names. Services are numbered rather than named in this table at the request of one participant, whose objection to a named comparison is printed in our correspondence; the methods are printed in full because the methods are the finding. Within-laboratory agreement was 0.1 to 0.3 points. Between-laboratory spread was 2.1 points, entirely accounted for by the disclosed method differences. | |||||
Four limitations, stated because the alternative is letting readers over-read a small study. First, one lot of one compound from one supplier is not a sample from which the performance of these services in general can be inferred; it is an existence proof about method-driven spread. Second, three services is too few for any statistical treatment beyond the descriptive; published round-robin studies of peptide purity use seven or more participants for exactly that reason.6
Third, and most important, the exercise tested reproducibility, not accuracy. All three could be equally wrong: without a certified reference standard of known purity, there is no true value against which to score them, and the compendial approach to validating a purity procedure requires exactly such a reference to establish accuracy rather than mere agreement.7 What we measured is dispersion around an unknown centre, and the same constraint applies to any quantitation attempted without a matched standard.8
Fourth, blind submission tests a laboratory’s ordinary process, which is the point, but it also means we bought the cheapest standard product from each service rather than the most thorough. A comparison of each service’s best available package would be a different and probably more flattering study, and it would tell a buyer less, because almost nobody buys the best available package.
The Journal will repeat the exercise annually with a different compound and, funding permitting, against a certified reference standard. The design is published so that others can run it.
A public, searchable archive of prior results changes the epistemics of this market more than any improvement in analytical method would. The reason is the selection problem. A single report is a snapshot chosen by whoever paid for it. A record of every submission relating to a vendor, published whether or not the result flattered the submitter, is a series — and a series cannot be curated by choosing which snapshots to show.
Two of the four services maintain something of this kind. The Journal would like all four to, and has asked. The principal objection raised in response is a real one: client confidentiality is an obligation, not a preference, and a laboratory cannot unilaterally publish a paying client’s results. We accept that entirely, which is why our request is narrower. Publish the fact of the submission — the date, the vendor named by the submitter, and the compound — while keeping the result confidential where the client requires it.
That narrower version defeats most of the selection effect at no cost to confidentiality. A vendor with nine public submissions and three published results is visibly a vendor with six results it declined to publish, and a reader can price that in. It is, so far as we can tell, the single highest-leverage change available to this sector, and it requires no new equipment and no new method.
Janoshik Analytical and PeptideMeter both advertise in The Compound Journal. Both relationships are disclosed by name on our funding page, together with every other sponsor. No advertiser sees editorial copy before publication, no advertiser has any role in commissioning or reviewing coverage, and the analytical-chemistry desk is contractually barred from consulting for any vendor, testing service or compounding pharmacy. This article was edited by the standards desk under the same rules as every other piece in the department.
The Journal also pays these services. We have submitted samples to three of the four on commercial terms, at list prices, and the blind duplicate exercise described above was funded from editorial budget. We are therefore simultaneously a customer of the institutions we are reporting on and a recipient of advertising revenue from two of them. Readers are entitled to weigh that, and the only useful response we can offer is to state it plainly and to publish objections.
Our position on the substance is unchanged by any of it. All four services are legitimate operations and we have no evidence of dishonesty by any of them. The problems this article describes are structural — who commissions testing, who decides what is published, and what a sample can support about a batch — and they would persist unchanged if every person working at all four organisations were beyond reproach. Correspondence to standards@compoundjournal.com.
A summary judgement, since a critical article of this length invites the inference that we think the sector is worthless. We do not. Independent testing in this market is the only mechanism by which a buyer can obtain information about material that is not supplied by the party selling it, and its existence is the difference between a market with some evidence in it and a market with none. Several of the reports these services produce are better documents than the manufacturer certificates they are checking, which is a low bar cleared with room to spare.
The three criticisms we would press are narrow. A sample is not a batch, and the trade cites samples as batches. The party paying for a test decides whether anybody sees it, and the visible corpus is therefore selected. And a badge on a listing has dropped every particular a reader would need. None of these is an analytical failing and none is a failing of the services in isolation; the second and third are properties of the market that surrounds them.
What we would tell a reader is this. A third-party report on a vial you selected and posted yourself is strong evidence about that vial. A third-party report published by the vendor is weaker evidence, of an amount you cannot determine. A badge is not evidence. And nothing in any of the three is a statement about whether anybody should administer the contents to anything.
We will repeat the blind duplicate exercise annually, with a different compound each year and, if we can fund it, against a certified reference standard so that accuracy rather than merely reproducibility can be assessed. The design is published in full so that anybody else can run it, and we would rather be contradicted by a better study than be the only publication that has tried.
The result is unremarkable. What the report omits is not.
The Journal submitted split samples from single lots to three assay services, under names unconnected to this publication, and published each method alongside each result.
The route did not close because of a rule about peptides.
A reminder that a purity figure is the output of a method, and that methods differ.
Follow the resin, not the catalogue.
Reported from the analysis, not from a warning notice.