Set the threshold at a tenth of a per cent and the small impurities vanish
We work through a single chromatogram twice, under two integration conventions, and show where the difference comes from.
TheCompound Journal
Reporting on incretins, compounding & the peptide supply chain
Method
A chromatogram is silent on everything that does not absorb: counter-ions, water, inorganic salts and non-chromophoric excipients all contribute mass to the vial and nothing to the trace.
What the Journal would like to see, and has seen from exactly one laboratory, is a report presenting two chromatograms from two orthogonal separations, stating both figures plainly, and reporting the lower of them as the result. That convention is conservative, it is transparent, and it is slightly commercially uncomfortable for whichever supplier commissioned the work, which is presumably why it has not spread. It is nonetheless the only version of a purity claim that has survived a deliberate attempt to falsify itself, and a claim that has survived such an attempt is a different kind of object from a claim that has never been tested at all.
The standard peptide mobile phase is water and acetonitrile with an acidic additive, and the choice of additive has consequences beyond pH. Trifluoroacetic acid at around a tenth of a per cent is the classical choice because it is an effective ion-pairing agent: it associates with basic residues, masks their charge, and produces markedly sharper and better-retained peaks than a simple pH adjustment achieves. For a difficult peptide separation the improvement is substantial.
It has two costs. Trifluoroacetate suppresses ionisation in electrospray mass spectrometry, which means a method optimised for chromatographic performance is frequently unsuitable for identity confirmation on the same run; formic acid is the usual compromise, giving worse peak shape and a usable mass spectrum. And residual trifluoroacetate from purification persists as a counter-ion, contributing mass to the vial and confounding content calculations, which is a manufacturing rather than an analytical issue but originates in the same chemistry.
Acetonitrile is preferred over methanol for peptide work on two grounds: lower viscosity, which means lower backpressure at a given flow, and lower ultraviolet absorbance at the short wavelengths peptide detection requires. Methanol’s absorbance in that region raises the baseline and degrades the signal-to-noise ratio exactly where the small impurity peaks are. A method reading at two hundred and fourteen nanometres in a methanol gradient is fighting its own solvent.
Integration software applies a threshold — expressed as a slope sensitivity, an area cut-off, a height cut-off or some combination — below which a feature in the trace is treated as baseline noise and not integrated. The setting is necessary: without it, every fluctuation would be reported as an impurity and the result would be dominated by noise. The setting is also consequential, because a great many real, small, closely related species live in the region between the two conventional choices.
The arithmetic is easy to underestimate. A well-made peptide preparation may carry twenty or thirty related species each between two-hundredths and a tenth of one per cent — deletion sequences, deamidated and oxidised forms, epimers. Reported individually against a low threshold they might total a percentage point or more. Discarded against a high threshold they total zero. Two laboratories reporting 99.4 and 98.3 on the same lot may have measured the same chromatogram and disagreed only about which features are noise.
The Journal has asked all four independent services what threshold their standard peptide report uses. Two answered with a figure. One answered that it depends on the method and offered to supply the value per report, which is a better answer than a fixed number. One did not answer. We regard the threshold as second only to the gradient in importance and, like the gradient, it is a single value that whoever produced the document already knows.
A method reading at 280 nanometres cannot see a fragment that has lost its aromatic residue, however much of it is there.
On detection wavelengthElectrospray ionisation of a peptide produces multiply charged ions, and the observed mass-to-charge series is deconvoluted to a molecular mass. Agreement with the theoretical mass of the intended sequence, within the accuracy of the instrument, is strong evidence that the molecule has the right elemental composition. It is not evidence that it has the right sequence, because permutations of the same residues have identical mass, and it is not evidence against isomeric degradation, because an isoaspartate rearrangement changes nothing about the mass.
Fragmentation closes most of that gap. Collision-induced dissociation of the peptide backbone produces a ladder of fragment ions whose mass differences read out the sequence, and a full or near-full ladder is genuine sequence confirmation. It requires a tandem instrument, more analyst time and a method that does not use an ionisation-suppressing additive, which is why identity work often runs on a formic acid gradient rather than the trifluoroacetic acid method used for purity.
The practical reading of a certificate follows. Identity confirmed by mass means the elemental composition matches. Identity confirmed by tandem mass spectrometry with sequence coverage means considerably more. Identity confirmed by retention-time comparison means the sample behaves like the standard. Three quite different claims are routinely expressed by the same phrase, and the difference between them is exactly the difference between knowing what is in the vial and knowing that it resembles something.1
| Laboratory | Purity | Gradient disclosed | Wavelength | Threshold | Chromatogram supplied |
|---|---|---|---|---|---|
| W | 99.1% | Run time only | 220 nm | Not stated | Yes |
| X | 98.5% | Full programme | 214 nm | 0.10% | Yes |
| Y | 97.6% | Full programme | 214 nm | 0.05% | Yes, two |
| Z | 98.8% | Not stated | Not stated | Not stated | No |
| Eight vials from a single lot, submitted in pairs, with no laboratory told the material was shared. Identities are withheld: none of the four agreed to be ranked, and what this table records is what reached the report rather than how well the analysis was done. Laboratory Y separated the sample twice on different principles and put the lower of its two figures on the front page, which is the cautious way to do it and the only instance we encountered. | |||||
Orthogonality is not a synonym for repetition. Two runs of the same method differ only in random variation. A shorter and a longer gradient on the same column separate by the same mechanism, and a pair of species co-eluting under one has a good chance of co-eluting under the other. Genuine orthogonality requires a different physical basis for the separation.
For peptides the practical options are well established. Changing mobile-phase pH alters the ionisation state of acidic and basic residues and therefore their effective hydrophobicity, frequently reordering closely eluting species — a peptide method at low pH and the same peptide at neutral pH are substantially different separations. Changing stationary-phase chemistry from octadecyl to phenyl or a polar-embedded phase alters selectivity by mechanism. Hydrophilic interaction chromatography inverts the retention principle. Ion-exchange separates by charge, and capillary electrophoresis by charge-to-size ratio in free solution.
The cost of a second method is instrument time on a sample already in the autosampler, and its value is that it can falsify the first result. Where the two agree, confidence rises substantially. Where they disagree, something is co-eluting and the lower figure is the safer one to report. One laboratory in this market runs two gradients as standard and reports the lower of the two figures; the Journal regards that as the single best analytical practice we have encountered in this trade, and it costs perhaps twenty minutes.2
First, how much peptide is in the vial. Counter-ions, residual water, inorganic salts and non-absorbing excipients contribute mass and no chromatographic signal, which is how a preparation can be 99 per cent pure and substantially less than 99 per cent peptide. Purity and content are different quantities and the second is the one that enters any calculation involving a mass.
Second, whether anything is aggregated. Reversed-phase conditions dissociate most non-covalent aggregates before detection, so the monomer is what arrives at the detector. Only a size-based separation reports high molecular weight species.
Third, whether the sequence is correct. Retention-time agreement is consistency; molecular mass is composition; only fragmentation approaches sequence. Fourth, whether an isomeric degradation product is present, since isoaspartate and racemised residues change nothing about mass and may or may not resolve depending on the method. Fifth, anything at all about microbiological quality — bioburden, sterility, endotoxin — which is a separate discipline in a separate laboratory.
Stated as a list it reads like an indictment of the technique, and it is not. Reversed-phase chromatography answers its own question superbly. The list is an indictment of a market that asks it five questions and prints one answer.
Across the twenty companies the Journal tracks, every certificate states a purity figure. Rather fewer state the method in any form. A minority name the detection wavelength. Almost none states the gradient as a rate or a programme, and we have seen an integration threshold on a supplier-issued certificate twice.
Practice that deserves naming: CPC states the wavelength and the run time on its standard certificate. SSA lists the three largest impurities with relative retention times, which is the single most informative addition we have seen anybody make. QST and BCH supplied full method parameters on request within a working day. WXT and SWB attach the third-party laboratory report rather than transcribing a figure, which removes a transcription step and with it a class of error. QYB, MKM, HJ, KP, SGN, FGP, ERP and JEEP follow the standard convention of a figure without a method, as do WWB, QSC, GGPeps, GL Biochem, Homopeptide and TFC.
The criticism is of the convention, which the whole market adopted collectively and any member of it could leave unilaterally. No company named here has been shown to us to have misstated a result, and where we have queried a figure against a chromatogram the responses have generally been prompt and technical. What we are describing is a document format that omits the four values needed to compare one number with another, and that omission is not in anybody’s interest, including the sellers’.
The compendial and regulatory material in this piece is taken from the current general chapters on chromatography and on validation of compendial procedures, from the European Pharmacopoeia chapters on liquid chromatography and on chromatographic separation techniques, and from the harmonised guidelines on analytical validation, on impurities and on specifications for biotechnological products, all read in the original. The separation science is drawn from the chromatography literature, with the peptide-specific behaviour cited where it differs from small-molecule practice.
Where the Journal reports a number it obtained, it states the number of vials, the number of laboratories, whether the vials came from one lot, whether the laboratories knew, and what method parameters were disclosed to us. Where we quote a figure from a certificate we state whether the method was disclosed on it. Where a laboratory or a company answered our questions we distinguish an answer from a refusal and a refusal from a non-response.
Nothing in this department is a recommendation to buy, use or avoid anything. The compounds referred to are sold for research use only and are not approved for human use in any jurisdiction. Corrections and disputes go to standards@compoundjournal.com. Readers with certificates or chromatograms they would like read should write to letters@compoundjournal.com; we do not identify the source of anything sent to us, and we do not publish a reader’s name without permission.
Readers who take one habit from this piece should take the second method. A single separation cannot detect its own co-elution, and a second run on a different principle costs instrument time on a sample already in the autosampler. Where two orthogonal figures agree, a purity claim has survived an attempt to break it. Where they disagree, the lower number is the one to write down.
We work through a single chromatogram twice, under two integration conventions, and show where the difference comes from.
Duplicate submissions under different names test within-laboratory repeatability, which is a different quantity from between-laboratory reproducibility.
Every third-party report in this market describes a sample somebody chose to send. That choice is outside the laboratory’s control and outside its records.
The condition on arrival is recorded by some services and not others, and it is one of the more informative lines in a report.
The supplier has not disputed the finding. It has not explained the gap either.
The Journal submitted split samples from single lots to three assay services, under names unconnected to this publication, and published each method alongside each result.