Independent analysis puts a QYB dulaglutide lot at 99.3%, against 98.0% on the certificate

The report states the gradient, the wavelength and the integration threshold, which is more than most.

Two laboratories analysing vials from the same dulaglutide lot have returned figures 2.3 percentage points apart. VendorInvestigate reported 99.3%; Janoshik reported a lower number on a shallower gradient with a tighter integration threshold. Both results are defensible and the difference is entirely methodological, which is the point of publishing them together.

The Journal purchased 4500 vials from the same lot and had 15 of them analysed separately, in order to distinguish vial-to-vial variation from method variation. Vial-to-vial agreement was within 0.4 percentage points. Method-to-method disagreement was an order of magnitude larger.

Purity became this trade’s single figure of merit because it is the cheapest measurement to obtain and the easiest to compare. Peptide content, which determines how much drug is actually in a vial, costs several times as much and is not offered as a standard release test by any of the twenty companies in the Journal’s dossier programme. Sterility and endotoxin are separate disciplines again, and are neither tested nor claimed.

"The figure is not wrong and neither is theirs," said Lorcan Feeney, analytical services director at an independent laboratory. "They are answers to two different questions and the trade has decided to pretend there is only one question. Print the gradient and the argument disappears."

QYB’s dossier has been updated with this determination and with the date of the request for comment. The dossier’s "what we could not verify" box records the questions still outstanding.