Twenty-eight days is a number from somebody else’s product

Every degradation pathway accelerates by orders of magnitude on reconstitution, because the solvent that lyophilisation removed is the reagent most of them need.

Reconstitution is the point at which a stable product becomes a perishable one, and the change is not incremental. Deamidation, hydrolysis and disulphide exchange all require water as a participant or a medium; aggregation requires molecular mobility that the dried matrix denies. Adding two millilitres of diluent restores all of it at once. The consequence is that a product with a two-year shelf life as a cake may have an in-use period measured in weeks, and the second figure is not derivable from the first by any calculation.

Residual moisture, and how it is measured

Two methods dominate. Karl Fischer titration determines water specifically, by a stoichiometric reaction with iodine, and is the reference method; the coulometric variant works on the small sample masses a single vial provides. Loss on drying is simpler and less specific, measuring total volatile mass lost under defined heating, which for a formulation containing residual organic solvent overstates the water.

Typical release specifications for lyophilised peptides sit in the range of one to three per cent water by mass, with tighter limits where the molecule is particularly moisture-sensitive. The relationship to stability is not linear. Below roughly one per cent, further drying sometimes destabilises rather than helps, because a monolayer of water contributes to conformational stability in some solid-state systems. Above three per cent, deamidation and hydrolysis rates rise steeply and the glass transition falls towards ambient.

None of the twenty companies the Journal tracks reports residual moisture as a standard release test. Two will provide a figure on request. This is the omission we would most like to see closed, ahead of endotoxin and well ahead of anything else, for a straightforwardly practical reason: it is a cheap determination on a small sample, it is performed in any pharmaceutical analytical laboratory, and it predicts what the vial will be like in eighteen months better than the purity figure that is printed instead.

Light, and the exposure nobody records

Photostability has its own guideline, its own defined light source options and its own exposure requirement expressed in lux hours of visible light and watt hours per square metre of near ultraviolet. Products are tested in the immediate container, and where they fail, in the marketing pack, and where they fail again the label carries a protection instruction. The chemistry is real: tryptophan and tyrosine absorb in the near ultraviolet and photo-oxidise, and photolytic disulphide cleavage is well documented.

Nothing about light exposure is recorded anywhere in the research-peptide supply chain. Vials are frequently supplied in clear glass. Photographs for listings are taken under studio lighting. Parcels are opened on kitchen counters. A reconstituted vial may sit on a shelf under a window for weeks. The cumulative exposure is unknown and unknowable, and it is plausibly a larger contributor to degradation than the transit excursions that attract all the attention.

The Journal makes one narrow observation rather than a recommendation, because recommendations are not this publication’s business. Amber glass, or a secondary carton, costs a fraction of a cent per unit and removes an uncontrolled variable entirely. Several of the twenty companies we track already ship in amber vials; most do not, and the ones that do not have not, as far as we can establish, generated any data suggesting it does not matter.1

Mean kinetic temperature is never lower than the arithmetic mean, and the gap is largest exactly where the excursion was shortest and hottest.

In-use stability, and where the numbers come from

In-use stability is established by a dedicated study: the finished product reconstituted as intended, at the intended concentration, in the intended container, stored at the intended temperature, sampled at intervals, and analysed by stability-indicating methods for related substances and by a size-based method for aggregates. The output is a period, and the period belongs to that formulation in that container and to nothing else.

The in-use periods circulating in this market are not derived that way. They are, in the Journal’s experience of tracing them, borrowed from the labelling of marketed pen presentations, which are different formulations at different concentrations with different preservative systems in different primary containers. Marketed in-use periods for the incretin pens run from four weeks to eight depending on product and storage condition, and none of those figures transfers to a reconstituted research vial by any argument we can construct.

What can be said generally is directional rather than numerical. Degradation in solution proceeds orders of magnitude faster than in the cake. Lower temperature helps substantially. Repeated warming and cooling of an opened vial is worse than steady storage. Preservative-containing diluent addresses microbial growth and does nothing about chemical degradation. And in the absence of a study on the actual product, any specific number quoted for an in-use period is an assumption wearing a specification’s clothes.

One further loss is routinely mistaken for degradation. Peptides adsorb to glass and polymer surfaces, and the relationship runs the awkward way: the more dilute the solution, the larger the proportion a given surface area removes.2

What twenty companies document on stability
Documented itemCompanies reporting as standardOn requestNot available
Storage condition, lyophilised2000
Storage condition stated separately for reconstituted6311
Shelf life or retest interval1901
Residual moisture0218
Study conditions supporting the shelf life0119
In-use period from a study on that product0020
Compiled from the standard release documentation of twenty companies tracked by the Journal, supplemented by a written questionnaire sent twice, four weeks apart. On request denotes a documented instance of the item being supplied when asked. The final row is the one we would most like to be able to revise.

Bacteriostatic water, sterile water, and what each is for

Sterile water for injection contains water and nothing else. It is sterile when the container is opened and it has no capacity to remain so, and it supports the growth of any organism introduced subsequently. It is the appropriate diluent for a single-use presentation and the wrong one for anything intended to be entered more than once.

Bacteriostatic water for injection contains benzyl alcohol at nine parts per thousand. Benzyl alcohol inhibits microbial growth, which is what makes a multiple-dose presentation coherent, and it is important to be exact about what that means: a preservative suppresses the proliferation of organisms introduced during use. It does not sterilise a contaminated solution, it does not act instantly, and its effectiveness against a given organism is established by a specific compendial test rather than assumed.

Two further points get lost. Benzyl alcohol is not universally compatible; it has been implicated in the aggregation of certain protein formulations, and compatibility with a given peptide is a question for data rather than for convention. And a preservative system has its own stability: preservative content declines over an in-use period, which is one of the attributes a proper in-use study measures. A diluent choice is therefore a formulation decision with chemical consequences, not a matter of preference between two clear liquids.3

Freezing a solution is not storing it

Freezing a reconstituted vial to extend its life is a common inference and a poor one, for reasons that have nothing to do with temperature and everything to do with what happens during the phase change. As ice forms, solutes are excluded from the crystal lattice and concentrated into a shrinking unfrozen fraction. Local concentration, ionic strength and pH in that fraction can shift dramatically — buffer components crystallise at different points, and a phosphate buffer is notorious for a large pH excursion on freezing.

The ice-water interface is itself a denaturing surface, and interfacial area increases with the number of freeze-thaw cycles. Each cycle presents the peptide with a fresh opportunity to unfold at that interface and aggregate. This is why formulations intended for frozen storage contain cryoprotectants and why lyophilisation exists as a technique at all: the point of drying is to avoid keeping a peptide in a partially frozen aqueous system.

The Journal states the mechanism and declines the recommendation, as this department’s practice requires. What can be said without advising anybody is that freezing a reconstituted solution is a different chemical operation from freezing a dried cake, that its effects are formulation-dependent and not predictable from first principles, and that no in-use study we have seen in this market has examined it. A reader treating the freezer as a pause button is relying on an assumption nobody has tested for that product.

Three lines that would make a stability claim checkable

As with sterility, none of what follows requires a regulator, and all of it is already known to whoever released the batch. First, report residual moisture with the method and the specification, or state that it is not determined. Second, state whether the shelf life is supported by a study on this product and this formulation, and if so under what condition and for how long — a single line reading twelve months at minus twenty, ongoing, would be worth more than any number currently printed. Third, use the words retest and expiry correctly, and say which applies.

A fourth would be welcome and is harder: state the in-use period and whether it derives from a study on this product. We expect that answer to be no almost everywhere, and a stated no is more useful than a borrowed twenty-eight days, because it tells a reader that the figure they were about to rely on does not exist.

The four independent testing services cannot close this gap and it would be unfair to ask them to. Janoshik, Medutest, PeptideMeter and VendorInvestigate can measure a vial today with real competence, and several will measure the same lot again later, which is the nearest thing to stability data available downstream. What none of them can sell is time, and a stability programme is mostly time.

A note on method and sourcing

The regulatory framework in this article is taken from the harmonised guidelines on stability testing and on biotechnological products, read in the original, and from the current compendial chapters on storage definitions, distribution of temperature-sensitive products and stability in dispensing practice. The degradation chemistry is drawn from the peptide and pharmaceutical sciences literature, and where a claim is a generalisation across sequences this piece says so, because sequence dependence is the rule rather than the exception.

The shipment data is ours. Nine parcels, ordered at catalogue prices as ordinary customers, with calibrated loggers placed inside the insulated payload and sampling at five-minute intervals. Eight complete traces and one truncated by a customs hold. We disclose that nine parcels is not a survey, that we did not control the packing operation, and that a single logger cannot characterise a payload with a thermal gradient across it.

Nothing in this department is a recommendation about storing, reconstituting or administering anything. The compounds discussed are sold for research use only and are not approved for human use in any jurisdiction. Corrections and disputes go to standards@compoundjournal.com; documents, traces and certificates readers would like examined go to letters@compoundjournal.com, and we do not identify the source of anything sent to us.

The finding the Journal least wanted and most trusts is that the transit excursions everybody worries about are probably not the largest exposure in this supply chain. A dried cake at low moisture tolerates a warm afternoon. A reconstituted vial in a refrigerator door for six weeks, with no in-use study behind the number that justified the six weeks, is a different proposition, and it attracts almost no attention at all.

References

  1. International Council for Harmonisation. Q1B: Photostability Testing of New Drug Substances and Products. 1996.
  2. “Surface adsorption losses of peptides at low concentration in glass and polymer containers.” Journal of Pharmaceutical Sciences. 2016;105(9):2617–2626.
  3. United States Pharmacopeia. General Chapter ⟨51⟩ Antimicrobial Effectiveness Testing. USP–NF, Rockville, MD.

Letters to the Editor

5 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

The mean kinetic temperature explanation is the clearest I have read anywhere, including in the training my employer paid for. I have printed the sidebar and put it on the wall of the dispatch room.

T. Elorriaga, San Sebastián

Your table of degradation pathways lists racemisation and then says it is essentially never reported. If it is never reported, on what basis do you list it as a real risk rather than a theoretical one?

N. Halvorsen, Trondheim

The Journal replies

On the basis of the synthesis and analytical literature, where epimer formation during solid-phase assembly and during storage at extremes of pH is well characterised. What is missing is not evidence that it occurs but evidence about how much of it is present in any particular commercial vial, which is a different absence and the one we should have named.

Parcel 7 reached thirty-eight degrees for nearly four hours and you then tell readers not to worry unduly. I accept the solid-state argument. I would still like to know what the material looked like on analysis, and your article does not say.

H. Baptiste, Fort-de-France

The Journal replies

A fair criticism of the reporting. Parcel 7 was submitted for purity determination on arrival and returned a figure within a percentage point of the supplier’s stated value, which is consistent with the solid-state argument and proves very little on its own, since we had no pre-shipment measurement on that vial. The design fault is ours: a shipment study without a paired baseline sample cannot answer the question we most wanted answered, and the next round will.

On amber glass: it is not merely cheap, it is standard in the wider chemical supply trade for anything with a chromophore. The fact that this market ships peptides in clear glass is a sign of who is doing the filling more than of any decision about photostability.

R. Hollenbeck, Spokane, WA

You draw a distinction between retest date and expiry date and then say suppliers use the wrong word. Which word do you think they should use, given that most of them have no study behind either?

J. Vasilenko, Chisinau

The Journal replies

Retest, with a stated interval and a note that no formal stability study supports it. That is an honest description of a chemical supplier’s position and it is standard practice in the wider chemical trade. Printing expiry implies a study exists, which is the specific inference we object to.

Related coverage

Explainers

One method is one opinion

An orthogonal method separates on a different physical principle, so that species co-eluting in the first are likely to resolve in the second. Two runs of the same method at…

Priya Ramanathan, PharmD·19 Dec 2024·9 min