Vol. 3, No. 6 — June 2026Independent since 2024

TheCompound Journal

Reporting on incretins, compounding & the peptide supply chain

A monthly journal of record.
30 issues · 32 contributors
Not medical advice. We sell nothing.

Purity

Mass spectrometry answers a different question, properly

Where two methods disagree, the conservative convention is to report the lower figure. It is not universal, and whether a laboratory follows it belongs on the report.

The reason it matters is that co-elution is the failure mode a single method cannot detect. A chromatogram showing one sharp, symmetrical peak and a purity of 99.3 per cent is consistent with a clean preparation and equally consistent with a preparation in which a related species elutes under the parent. Peak-purity assessment from a diode-array detector helps and is not conclusive. A second separation on a different principle is the practical answer, and it is cheap: the sample is already in the vial, the second run costs instrument time.

Mass spectrometry, and the limits of a matching mass

Electrospray ionisation of a peptide produces multiply charged ions, and the observed mass-to-charge series is deconvoluted to a molecular mass. Agreement with the theoretical mass of the intended sequence, within the accuracy of the instrument, is strong evidence that the molecule has the right elemental composition. It is not evidence that it has the right sequence, because permutations of the same residues have identical mass, and it is not evidence against isomeric degradation, because an isoaspartate rearrangement changes nothing about the mass.

Fragmentation closes most of that gap. Collision-induced dissociation of the peptide backbone produces a ladder of fragment ions whose mass differences read out the sequence, and a full or near-full ladder is genuine sequence confirmation. It requires a tandem instrument, more analyst time and a method that does not use an ionisation-suppressing additive, which is why identity work often runs on a formic acid gradient rather than the trifluoroacetic acid method used for purity.

The practical reading of a certificate follows. Identity confirmed by mass means the elemental composition matches. Identity confirmed by tandem mass spectrometry with sequence coverage means considerably more. Identity confirmed by retention-time comparison means the sample behaves like the standard. Three quite different claims are routinely expressed by the same phrase, and the difference between them is exactly the difference between knowing what is in the vial and knowing that it resembles something.1

The second method, and what makes it orthogonal

Orthogonality is not a synonym for repetition. Two runs of the same method differ only in random variation. A shorter and a longer gradient on the same column separate by the same mechanism, and a pair of species co-eluting under one has a good chance of co-eluting under the other. Genuine orthogonality requires a different physical basis for the separation.

For peptides the practical options are well established. Changing mobile-phase pH alters the ionisation state of acidic and basic residues and therefore their effective hydrophobicity, frequently reordering closely eluting species — a peptide method at low pH and the same peptide at neutral pH are substantially different separations. Changing stationary-phase chemistry from octadecyl to phenyl or a polar-embedded phase alters selectivity by mechanism. Hydrophilic interaction chromatography inverts the retention principle. Ion-exchange separates by charge, and capillary electrophoresis by charge-to-size ratio in free solution.

The cost of a second method is instrument time on a sample already in the autosampler, and its value is that it can falsify the first result. Where the two agree, confidence rises substantially. Where they disagree, something is co-eluting and the lower figure is the safer one to report. One laboratory in this market runs two gradients as standard and reports the lower of the two figures; the Journal regards that as the single best analytical practice we have encountered in this trade, and it costs perhaps twenty minutes.2

Reversed-phase chromatography answers its own question superbly. The market asks it five questions and prints one answer.

Orla McCaffrey, Staff Writer, Analytics

Five things a purity figure cannot tell you

First, how much peptide is in the vial. Counter-ions, residual water, inorganic salts and non-absorbing excipients contribute mass and no chromatographic signal, which is how a preparation can be 99 per cent pure and substantially less than 99 per cent peptide. Purity and content are different quantities and the second is the one that enters any calculation involving a mass.

Second, whether anything is aggregated. Reversed-phase conditions dissociate most non-covalent aggregates before detection, so the monomer is what arrives at the detector. Only a size-based separation reports high molecular weight species.

Third, whether the sequence is correct. Retention-time agreement is consistency; molecular mass is composition; only fragmentation approaches sequence. Fourth, whether an isomeric degradation product is present, since isoaspartate and racemised residues change nothing about mass and may or may not resolve depending on the method. Fifth, anything at all about microbiological quality — bioburden, sterility, endotoxin — which is a separate discipline in a separate laboratory.

Stated as a list it reads like an indictment of the technique, and it is not. Reversed-phase chromatography answers its own question superbly. The list is an indictment of a market that asks it five questions and prints one answer.

Detection wavelength and what responds to it
WavelengthPrincipal absorberSees fragments without aromaticsBaseline noiseTypical use
214 nmAmide bondYesHigherPeptide purity and related substances
220 nmAmide bondYesModeratePeptide purity, quieter baseline
254 nmAromatic systemsNoLowSmall-molecule work, legacy detectors
280 nmTrp, Tyr, Phe side chainsNoLowTracking an aromatic residue; not a purity method
Diode array, 200–400 nmAll of the aboveYesMethod-dependentPeak purity assessment, spectral homogeneity
A purity figure generated at 280 nm is not comparable with one generated at 214 nm, and the difference is not a matter of a percentage point. Certificates reading at 280 nm circulate in this market; readers have sent us several.

A note on method and sourcing

The compendial and regulatory material in this piece is taken from the current general chapters on chromatography and on validation of compendial procedures, from the European Pharmacopoeia chapters on liquid chromatography and on chromatographic separation techniques, and from the harmonised guidelines on analytical validation, on impurities and on specifications for biotechnological products, all read in the original. The separation science is drawn from the chromatography literature, with the peptide-specific behaviour cited where it differs from small-molecule practice.

Where the Journal reports a number it obtained, it states the number of vials, the number of laboratories, whether the vials came from one lot, whether the laboratories knew, and what method parameters were disclosed to us. Where we quote a figure from a certificate we state whether the method was disclosed on it. Where a laboratory or a company answered our questions we distinguish an answer from a refusal and a refusal from a non-response.

Nothing in this department is a recommendation to buy, use or avoid anything. The compounds referred to are sold for research use only and are not approved for human use in any jurisdiction. Corrections and disputes go to standards@compoundjournal.com. Readers with certificates or chromatograms they would like read should write to letters@compoundjournal.com; we do not identify the source of anything sent to us, and we do not publish a reader’s name without permission.

The full spectrum, recorded and discarded

A diode-array detector records ultraviolet absorbance across a wavelength range—commonly 200 to 400 nanometres—at every point in the chromatogram. From that data, two things of immediate value can be extracted. First, the spectrum can be examined at each peak to confirm spectral homogeneity—a pure peak has a spectrum that does not drift across its width, while a co-eluted pair shows a spectrum that changes. Second, the full spectrum can be plotted to reveal absorbers that were not visible at the single monitoring wavelength.3 A 254-nanometre impurity in a sample monitored at 280 nanometres is invisible in the 214-nanometre chromatogram and is therefore invisible in the purity report, until a diode-array scan reveals it.

Almost no research-peptide certificate prints this information. The data exists—it is generated automatically—and its absence from the report is a formatting decision, not a technical limitation. The usefulness of seeing the full spectrum would be apparent on the first certificate where it resolves an otherwise inexplicable discrepancy between the purity and the mass-balance calculation. It is one of the few changes to the standard format that costs nothing and adds information the reader genuinely needs.

Readers who take one habit from this piece should take the second method. A single separation cannot detect its own co-elution, and a second run on a different principle costs instrument time on a sample already in the autosampler. Where two orthogonal figures agree, a purity claim has survived an attempt to break it. Where they disagree, the lower number is the one to write down.

References

  1. “Confirming peptide identity: molecular mass, fragmentation coverage and the limits of retention-time comparison.” Journal of Peptide Science. 2019;25(8):e3195.
  2. “Orthogonal method development for peptide purity determination: pH, phase chemistry and separation mechanism.” Journal of Chromatography A. 2020;1618:460873.
  3. European Pharmacopoeia. Chapter 2.2.29 — Liquid Chromatography. Council of Europe, Strasbourg.

Letters to the Editor

5 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

Your worked example varies gradient and threshold together and reports a 1.8-point spread. Which of the two contributed more? The article does not say, and the answer matters for what you are asking suppliers to disclose first.

M. Karlsen, Kristiansand

The Journal replies

Gradient, by roughly two to one in our four conditions: holding the threshold at 0.10 per cent, lengthening the gradient cost 0.9 points, while holding the gradient and tightening the threshold cost 0.4 to 0.9 depending on which gradient. We should have printed that decomposition in the table and it now appears in the note. If a supplier will disclose only one value, it should be the gradient.

Something your article omits, and it changes where the responsibility sits. Method selection is frequently specified by the customer, not by us. A purchase order arrives asking for a peptide purity run at a stated price and turnaround, and the method that fits those two constraints is the method that runs. We are perfectly willing to develop a longer separation for anybody who wants one, and in eleven years almost nobody has asked.

N. Fairweather, Hamilton

The Journal replies

That is a genuinely different account of the causation from the one we gave, and if it generalises it matters. Our piece treats method choice as a laboratory decision and yours treats it as a procurement decision. We would like to test which it is, and we are writing to the four independent services to ask what proportion of incoming work specifies a method at all.

You write that only one laboratory attached its chromatogram to the private buyer report. That was probably us. We started doing it five years ago because the PDF seemed incomplete without it. It costs us nothing to add — the instrument generates it automatically — and it solves exactly the dispute-resolution problem you describe. More laboratories should do it, and the reason they do not is not technical.

G. Papadakis, Thessaloniki

The Journal replies

That is generous of you to say. The technical barrier is near zero, and if enough laboratories began printing them, it would force the convention to change across the market. It is an example of something that costs one actor almost nothing but creates value for everyone, and it is precisely the kind of thing that can shift a trade practice when a few leaders move first.

On the section about diode-array detection and peak purity, I would add that true peak purity assessment requires library matching or at least spectral comparison across the peak width. A homogeneous spectrum tells you the peak is probably pure. A spectrum that shifts across the peak tells you it is not, and that information closes a gap the article identifies correctly.

O. Brannigan, Galway

The table showing what each method can detect is valuable but incomplete. You show no row for C-terminal truncation or N-terminal truncation as distinct phenomena. These are not rare, and they often elute differently depending on which end is missing. A generic gradient might resolve them; an improperly designed orthogonal method might not. The capability matrix should separate these cases.

C. Wilcoxson, Des Moines, IA

Related coverage