What changed at TFC in December, and what the company will not say about it
Documentation practice is the only part of vendor quality a buyer can assess before purchase.
TheCompound Journal
Reporting on incretins, compounding & the peptide supply chain
Purity
Where two methods disagree, the conservative convention is to report the lower figure. It is not universal, and whether a laboratory follows it belongs on the report.
The reason it matters is that co-elution is the failure mode a single method cannot detect. A chromatogram showing one sharp, symmetrical peak and a purity of 99.3 per cent is consistent with a clean preparation and equally consistent with a preparation in which a related species elutes under the parent. Peak-purity assessment from a diode-array detector helps and is not conclusive. A second separation on a different principle is the practical answer, and it is cheap: the sample is already in the vial, the second run costs instrument time.
Purity won on economics. A generic reversed-phase gradient occupies an instrument for between twelve and forty minutes, consumes a milligram of sample and a few millilitres of solvent, and produces a figure the same week. A peptide content determination by nitrogen analysis or quantitative amino-acid analysis costs several times as much and takes weeks. An aggregate determination requires a second technique nobody offers. Endotoxin requires a different laboratory. Sterility requires a fortnight and destroys the container.
Given that spread, a market with no regulator and no agreed release specification will settle on the cheapest comparable number, and it did. The trouble is what happens next. Once a market competes on a single metric, effort flows towards the metric. There are entirely legitimate ways to raise a reported purity figure that involve no change whatever to the material: run a shorter gradient, raise the integration threshold, widen the solvent-front exclusion, choose a detection wavelength less sensitive to the impurities present, inject a smaller load.
None of those is fraud. Each is a defensible analytical choice with a published rationale. Collectively they mean that the difference between a 99.4 and a 97.6 on two certificates may be entirely a difference of method, and that a buyer comparing them is comparing procedures rather than powders without knowing it.1
The amide bond has a strong absorbance in the far ultraviolet, and peptide methods conventionally read at two hundred and fourteen or two hundred and twenty nanometres to exploit it. The virtue is universality: every peptide-bonded species responds, roughly in proportion to the number of bonds it contains, which is as close to a mass-proportional response as ultraviolet detection gets. The cost is that solvents, additives and dissolved gases also absorb there, so baseline noise is higher and mobile-phase quality matters more.
Aromatic side chains absorb near two hundred and eighty nanometres, where the backbone is essentially transparent. A method reading there sees only species containing tryptophan, tyrosine or phenylalanine, on a quiet baseline. For a peptide with a single tryptophan it is a selective and elegant way to track that residue. As a purity method it is close to indefensible, because any fragment that has lost the aromatic residue is invisible regardless of how much is present.
Certificates reading at two hundred and eighty nanometres do circulate. Readers have sent us several. The Journal’s position is not that such a method is wrong but that it answers a different question, that a purity figure derived from it is not comparable with one derived at two hundred and fourteen, and that the wavelength is one line and belongs on the page. A diode-array detector records everything at once and makes the entire argument moot, which is why we ask whether one was used.
A method reading at 280 nanometres cannot see a fragment that has lost its aromatic residue, however much of it is there.
On detection wavelengthA detector responds linearly to concentration over a defined range and then stops. Overload the column or saturate the detector with too much sample and the main peak flattens at the top, its apex broadens, and its integrated area no longer represents the quantity present. Since the main peak is the numerator and dominates the denominator, distorting it distorts the purity figure — usually downwards, because the flattened peak loses area relative to a properly loaded one.
There is a competing pressure, and it is the reason overloading happens. Small impurities near the reporting threshold need adequate signal-to-noise to be integrated at all, and the way to raise their signal is to inject more sample. A laboratory hunting for 0.05 per cent impurities is tempted towards a load that compromises the main peak. The correct answer in regulated practice is two injections: a small load for the main peak and a larger one for the related-substances profile, with the results combined.
Column overload is a separate phenomenon from detector saturation and produces a characteristic asymmetric fronting peak. Both are visible on the chromatogram to anybody who is shown it, which is one of several reasons the Journal asks for the trace rather than the number. A purity figure calculated from a distorted main peak is arithmetically correct and analytically meaningless, and the only way to know is to look.2
| Laboratory | Purity | Gradient disclosed | Wavelength | Threshold | Chromatogram supplied |
|---|---|---|---|---|---|
| W | 99.1% | Run time only | 220 nm | Not stated | Yes |
| X | 98.5% | Full programme | 214 nm | 0.10% | Yes |
| Y | 97.6% | Full programme | 214 nm | 0.05% | Yes, two |
| Z | 98.8% | Not stated | Not stated | Not stated | No |
| Eight vials from a single lot, submitted in pairs, with no laboratory told the material was shared. Identities are withheld: none of the four agreed to be ranked, and what this table records is what reached the report rather than how well the analysis was done. Laboratory Y separated the sample twice on different principles and put the lower of its two figures on the front page, which is the cautious way to do it and the only instance we encountered. | |||||
Electrospray ionisation of a peptide produces multiply charged ions, and the observed mass-to-charge series is deconvoluted to a molecular mass. Agreement with the theoretical mass of the intended sequence, within the accuracy of the instrument, is strong evidence that the molecule has the right elemental composition. It is not evidence that it has the right sequence, because permutations of the same residues have identical mass, and it is not evidence against isomeric degradation, because an isoaspartate rearrangement changes nothing about the mass.
Fragmentation closes most of that gap. Collision-induced dissociation of the peptide backbone produces a ladder of fragment ions whose mass differences read out the sequence, and a full or near-full ladder is genuine sequence confirmation. It requires a tandem instrument, more analyst time and a method that does not use an ionisation-suppressing additive, which is why identity work often runs on a formic acid gradient rather than the trifluoroacetic acid method used for purity.
The practical reading of a certificate follows. Identity confirmed by mass means the elemental composition matches. Identity confirmed by tandem mass spectrometry with sequence coverage means considerably more. Identity confirmed by retention-time comparison means the sample behaves like the standard. Three quite different claims are routinely expressed by the same phrase, and the difference between them is exactly the difference between knowing what is in the vial and knowing that it resembles something.3
Orthogonality is not a synonym for repetition. Two runs of the same method differ only in random variation. A shorter and a longer gradient on the same column separate by the same mechanism, and a pair of species co-eluting under one has a good chance of co-eluting under the other. Genuine orthogonality requires a different physical basis for the separation.
For peptides the practical options are well established. Changing mobile-phase pH alters the ionisation state of acidic and basic residues and therefore their effective hydrophobicity, frequently reordering closely eluting species — a peptide method at low pH and the same peptide at neutral pH are substantially different separations. Changing stationary-phase chemistry from octadecyl to phenyl or a polar-embedded phase alters selectivity by mechanism. Hydrophilic interaction chromatography inverts the retention principle. Ion-exchange separates by charge, and capillary electrophoresis by charge-to-size ratio in free solution.
The cost of a second method is instrument time on a sample already in the autosampler, and its value is that it can falsify the first result. Where the two agree, confidence rises substantially. Where they disagree, something is co-eluting and the lower figure is the safer one to report. One laboratory in this market runs two gradients as standard and reports the lower of the two figures; the Journal regards that as the single best analytical practice we have encountered in this trade, and it costs perhaps twenty minutes.4
First, how much peptide is in the vial. Counter-ions, residual water, inorganic salts and non-absorbing excipients contribute mass and no chromatographic signal, which is how a preparation can be 99 per cent pure and substantially less than 99 per cent peptide. Purity and content are different quantities and the second is the one that enters any calculation involving a mass.
Second, whether anything is aggregated. Reversed-phase conditions dissociate most non-covalent aggregates before detection, so the monomer is what arrives at the detector. Only a size-based separation reports high molecular weight species.
Third, whether the sequence is correct. Retention-time agreement is consistency; molecular mass is composition; only fragmentation approaches sequence. Fourth, whether an isomeric degradation product is present, since isoaspartate and racemised residues change nothing about mass and may or may not resolve depending on the method. Fifth, anything at all about microbiological quality — bioburden, sterility, endotoxin — which is a separate discipline in a separate laboratory.
Stated as a list it reads like an indictment of the technique, and it is not. Reversed-phase chromatography answers its own question superbly. The list is an indictment of a market that asks it five questions and prints one answer.
The compendial and regulatory material in this piece is taken from the current general chapters on chromatography and on validation of compendial procedures, from the European Pharmacopoeia chapters on liquid chromatography and on chromatographic separation techniques, and from the harmonised guidelines on analytical validation, on impurities and on specifications for biotechnological products, all read in the original. The separation science is drawn from the chromatography literature, with the peptide-specific behaviour cited where it differs from small-molecule practice.
Where the Journal reports a number it obtained, it states the number of vials, the number of laboratories, whether the vials came from one lot, whether the laboratories knew, and what method parameters were disclosed to us. Where we quote a figure from a certificate we state whether the method was disclosed on it. Where a laboratory or a company answered our questions we distinguish an answer from a refusal and a refusal from a non-response.
Nothing in this department is a recommendation to buy, use or avoid anything. The compounds referred to are sold for research use only and are not approved for human use in any jurisdiction. Corrections and disputes go to standards@compoundjournal.com. Readers with certificates or chromatograms they would like read should write to letters@compoundjournal.com; we do not identify the source of anything sent to us, and we do not publish a reader’s name without permission.
Readers who take one habit from this piece should take the second method. A single separation cannot detect its own co-elution, and a second run on a different principle costs instrument time on a sample already in the autosampler. Where two orthogonal figures agree, a purity claim has survived an attempt to break it. Where they disagree, the lower number is the one to write down.
Documentation practice is the only part of vendor quality a buyer can assess before purchase.
Two years ago we ran an anonymised version of this comparison and promised a named one. This is it, with every method printed in full.
Nothing on this page is difficult. It is simply unfamiliar, and unfamiliarity is what makes a weak certificate look like a strong one.
System suitability is the set of checks demonstrating that the instrument and method were performing adequately when your sample was injected. It is recorded as a matter of…
A reminder that a purity figure is the output of a method, and that methods differ.
The report states the gradient, the wavelength and the integration threshold, which is more than most.