The conservative convention: report the lower of two numbers

An orthogonal method separates on a different physical principle, so that species co-eluting in the first are likely to resolve in the second. Two runs of the same method at different speeds are not orthogonal.

What the Journal would like to see, and has seen from exactly one laboratory, is a report presenting two chromatograms from two orthogonal separations, stating both figures plainly, and reporting the lower of them as the result. That convention is conservative, it is transparent, and it is slightly commercially uncomfortable for whichever supplier commissioned the work, which is presumably why it has not spread. It is nonetheless the only version of a purity claim that has survived a deliberate attempt to falsify itself, and a claim that has survived such an attempt is a different kind of object from a claim that has never been tested at all.

Reference standards, and what a retention time proves

A reference standard is material of established identity, purity and content against which an analysis is calibrated. Compendial standards are characterised by collaborative study and supplied with a certificate stating their assigned content. In-house standards are qualified against a compendial standard where one exists, or characterised by a battery of orthogonal methods where one does not. For most research peptides there is no compendial standard, which means every claim of identity or assay in this market ultimately rests on somebody’s in-house material.

What a matched retention time supports is worth stating precisely: it supports the inference that the sample and the standard behave identically in this separation. That is real evidence of consistency between two materials. It is not identification, because retention time is not unique — deletion sequences, epimers and unrelated compounds of similar hydrophobicity can share a retention window, and the peak width of a peptide separation is wide enough to hide a great deal.

Two consequences follow for reading a certificate. A report stating that identity was confirmed by comparison of retention time with a reference standard has told you about consistency, not identity. And a purity figure quoted as an assay — a percentage of label claim — requires a quantitative standard of known content, which is a much stronger claim than area per cent and should be labelled differently. The two are routinely printed in the same field.1

Mass spectrometry, and the limits of a matching mass

Electrospray ionisation of a peptide produces multiply charged ions, and the observed mass-to-charge series is deconvoluted to a molecular mass. Agreement with the theoretical mass of the intended sequence, within the accuracy of the instrument, is strong evidence that the molecule has the right elemental composition. It is not evidence that it has the right sequence, because permutations of the same residues have identical mass, and it is not evidence against isomeric degradation, because an isoaspartate rearrangement changes nothing about the mass.

Fragmentation closes most of that gap. Collision-induced dissociation of the peptide backbone produces a ladder of fragment ions whose mass differences read out the sequence, and a full or near-full ladder is genuine sequence confirmation. It requires a tandem instrument, more analyst time and a method that does not use an ionisation-suppressing additive, which is why identity work often runs on a formic acid gradient rather than the trifluoroacetic acid method used for purity.

The practical reading of a certificate follows. Identity confirmed by mass means the elemental composition matches. Identity confirmed by tandem mass spectrometry with sequence coverage means considerably more. Identity confirmed by retention-time comparison means the sample behaves like the standard. Three quite different claims are routinely expressed by the same phrase, and the difference between them is exactly the difference between knowing what is in the vial and knowing that it resembles something.2

Resolution is the joint product of efficiency and selectivity. Improvement in one does not compensate for inadequacy in the other.

On the method trade-off

The second method, and what makes it orthogonal

Orthogonality is not a synonym for repetition. Two runs of the same method differ only in random variation. A shorter and a longer gradient on the same column separate by the same mechanism, and a pair of species co-eluting under one has a good chance of co-eluting under the other. Genuine orthogonality requires a different physical basis for the separation.

For peptides the practical options are well established. Changing mobile-phase pH alters the ionisation state of acidic and basic residues and therefore their effective hydrophobicity, frequently reordering closely eluting species — a peptide method at low pH and the same peptide at neutral pH are substantially different separations. Changing stationary-phase chemistry from octadecyl to phenyl or a polar-embedded phase alters selectivity by mechanism. Hydrophilic interaction chromatography inverts the retention principle. Ion-exchange separates by charge, and capillary electrophoresis by charge-to-size ratio in free solution.

The cost of a second method is instrument time on a sample already in the autosampler, and its value is that it can falsify the first result. Where the two agree, confidence rises substantially. Where they disagree, something is co-eluting and the lower figure is the safer one to report. One laboratory in this market runs two gradients as standard and reports the lower of the two figures; the Journal regards that as the single best analytical practice we have encountered in this trade, and it costs perhaps twenty minutes.3

One sample, four method conditions, four purity figures
ConditionGradient rate (%ACN/min)Run time (min)ThresholdPurity reported
A1.67120.10%99.3%
B1.67120.05%98.9%
C0.50400.10%98.4%
D0.50400.05%97.5%
One physical sample from one vial, one instrument, one analyst, one afternoon; 25–45% acetonitrile in both gradients, 214 nm, identical column and injection load. The 1.8-point spread is attributable entirely to gradient slope and integration threshold. The conditions were specified by the Journal and do not represent the standard practice of the laboratory concerned.

Five things a purity figure cannot tell you

First, how much peptide is in the vial. Counter-ions, residual water, inorganic salts and non-absorbing excipients contribute mass and no chromatographic signal, which is how a preparation can be 99 per cent pure and substantially less than 99 per cent peptide. Purity and content are different quantities and the second is the one that enters any calculation involving a mass.

Second, whether anything is aggregated. Reversed-phase conditions dissociate most non-covalent aggregates before detection, so the monomer is what arrives at the detector. Only a size-based separation reports high molecular weight species.

Third, whether the sequence is correct. Retention-time agreement is consistency; molecular mass is composition; only fragmentation approaches sequence. Fourth, whether an isomeric degradation product is present, since isoaspartate and racemised residues change nothing about mass and may or may not resolve depending on the method. Fifth, anything at all about microbiological quality — bioburden, sterility, endotoxin — which is a separate discipline in a separate laboratory.

Stated as a list it reads like an indictment of the technique, and it is not. Reversed-phase chromatography answers its own question superbly. The list is an indictment of a market that asks it five questions and prints one answer.

Documentation practice, named

Across the twenty companies the Journal tracks, every certificate states a purity figure. Rather fewer state the method in any form. A minority name the detection wavelength. Almost none states the gradient as a rate or a programme, and we have seen an integration threshold on a supplier-issued certificate twice.

Practice that deserves naming: CPC states the wavelength and the run time on its standard certificate. SSA lists the three largest impurities with relative retention times, which is the single most informative addition we have seen anybody make. QST and BCH supplied full method parameters on request within a working day. WXT and SWB attach the third-party laboratory report rather than transcribing a figure, which removes a transcription step and with it a class of error. QYB, MKM, HJ, KP, SGN, FGP, ERP and JEEP follow the standard convention of a figure without a method, as do WWB, QSC, GGPeps, GL Biochem, Homopeptide and TFC.

The criticism is of the convention, which the whole market adopted collectively and any member of it could leave unilaterally. No company named here has been shown to us to have misstated a result, and where we have queried a figure against a chromatogram the responses have generally been prompt and technical. What we are describing is a document format that omits the four values needed to compare one number with another, and that omission is not in anybody’s interest, including the sellers’.

A note on method and sourcing

The compendial and regulatory material in this piece is taken from the current general chapters on chromatography and on validation of compendial procedures, from the European Pharmacopoeia chapters on liquid chromatography and on chromatographic separation techniques, and from the harmonised guidelines on analytical validation, on impurities and on specifications for biotechnological products, all read in the original. The separation science is drawn from the chromatography literature, with the peptide-specific behaviour cited where it differs from small-molecule practice.

Where the Journal reports a number it obtained, it states the number of vials, the number of laboratories, whether the vials came from one lot, whether the laboratories knew, and what method parameters were disclosed to us. Where we quote a figure from a certificate we state whether the method was disclosed on it. Where a laboratory or a company answered our questions we distinguish an answer from a refusal and a refusal from a non-response.

Nothing in this department is a recommendation to buy, use or avoid anything. The compounds referred to are sold for research use only and are not approved for human use in any jurisdiction. Corrections and disputes go to standards@compoundjournal.com. Readers with certificates or chromatograms they would like read should write to letters@compoundjournal.com; we do not identify the source of anything sent to us, and we do not publish a reader’s name without permission.

The technique is not on trial here and never was. Reversed-phase chromatography can resolve species differing by a single methyl group, and the laboratories running it for this market are, on the evidence we have gathered, largely competent and entirely willing to describe what they did when somebody asks. What is on trial is a document format that omits the four values needed to compare one figure with another.

References

  1. United States Pharmacopeia. General Chapter ⟨1225⟩ Validation of Compendial Procedures. USP–NF, Rockville, MD.
  2. “Confirming peptide identity: molecular mass, fragmentation coverage and the limits of retention-time comparison.” Journal of Peptide Science. 2019;25(8):e3195.
  3. “Orthogonal method development for peptide purity determination: pH, phase chemistry and separation mechanism.” Journal of Chromatography A. 2020;1618:460873.

Letters to the Editor

4 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

The table showing what each method can detect is valuable but incomplete. You show no row for C-terminal truncation or N-terminal truncation as distinct phenomena. These are not rare, and they often elute differently depending on which end is missing. A generic gradient might resolve them; an improperly designed orthogonal method might not. The capability matrix should separate these cases.

A. Fournier, Nantes

Your article argues for two orthogonal methods and reports the lower figure, but the people running a single twelve-minute method have a cost story you do not address. A full orthogonal pair doubles the turnaround and at least doubles the cost, which is why the market does not do it. The criticism of method disclosure is fair. The criticism that a single method is wrong is unfair to the constraints people operate under.

L. Fontaine, Brussels

The Journal replies

We are careful to say that a second method costs instrument time on a sample already in the autosampler, which is substantially less than twice the turnaround, but you are right that we underweight the commercial reality that a buyer setting a budget for testing is trading thoroughness for speed and price. Where we would push back is that those constraints are not technical or regulatory ones. They are market ones, and markets can change if enough buyers demand it.

Sub-two-micron columns changed everything about what is practical for peptide separations, but I would push back on the statement that particle size gains are costless. The pressure limit of most commercial instruments is three hundred bar, and trying to force 1.7-micron particles at eight millilitres per minute on a 4.6-millimetre column will send you there quickly. Peak capacity is not free.

P. Ekundayo, Akure

The Journal replies

Correct on the pressure cost, and that belongs in the method section. The trade-off is real, which is why many laboratories use core-shell superficially porous particles as a compromise: they are substantially cheaper than sub-two-micron packings, deliver most of the efficiency gain at a lower pressure, and the efficiency-cost-pressure triangle is the actual business decision people make. We should have named it.

One practical note from the other side of the counter. When a customer asks us for the gradient we send it. When a customer asks a reseller, the reseller does not have it, because they were sent a PDF with a number on it. The gap you are describing is often two links down the chain rather than at the laboratory.

J. Wenninger, Graz

The Journal replies

That is an important structural point and it changes where the fix has to happen. If the laboratory report travels intact instead of being transcribed into a house certificate, the method travels with it. Four of the twenty companies we track already attach the original report, and on this argument they are doing the single most useful thing available.