Vol. 3, No. 6 — June 2026Independent since 2024

TheCompound Journal

Reporting on incretins, compounding & the peptide supply chain

A monthly journal of record.
30 issues · 32 contributors
Not medical advice. We sell nothing.

Identity

Why a 0.036-dalton difference is the hardest number in peptide identity

Parts per million sound impressive until they are converted into daltons at the molecular weight of the thing being measured.

Parts per million are a flattering unit, and they need converting before they mean anything. Five parts per million on a peptide of four thousand daltons is a tolerance of two hundredths of a dalton, which is excellent and sufficient to exclude most compositional substitutions. One hundred parts per million on the same peptide is four tenths of a dalton: enough to confirm the compound is broadly the right one, not enough to exclude a deamidated form. Five hundred parts per million, an ordinary figure for a linear time-of-flight instrument at this mass, is two daltons, a tolerance inside which most of the interesting questions disappear.

The isotope pattern, and how it declares the charge

Because carbon-13 is present at roughly 1.1% natural abundance, a peptide containing one hundred and ninety carbon atoms will exist substantially as molecules containing one, two or three carbon-13 atoms. In a spectrum this appears as a series of peaks above the monoisotopic peak, separated in mass by approximately 1.00336 daltons and distributed in intensity according to the binomial statistics of the composition.

Two things follow, and both are practically useful. First, the spacing between adjacent isotope peaks in a charge-state cluster is one over the charge: a spacing of 0.5 on the m/z axis means the ion is doubly charged, 0.333 means triply, 0.25 means quadruply. This is the simplest charge assignment available and it requires no assumptions about the sample at all. Second, the relative intensities of the isotope peaks are predictable from the elemental composition, so a cluster whose shape departs markedly from the calculated envelope is evidence that two species are overlapping.

Both observations require an instrument capable of resolving the isotope peaks at the relevant m/z, which is where resolving power stops being a specification-sheet number and becomes the thing that determines whether a spectrum can be interpreted at all. Below roughly ten thousand resolving power, a multiply charged peptide envelope collapses into a single broad hump that carries neither the spacing nor the shape information.

Resolving power, defined and then converted

Resolving power is conventionally defined as m divided by Δm, where Δm is the width of the peak at half its maximum height. An instrument quoted at 30,000 resolving power at m/z 1000 produces peaks roughly 0.033 wide at that position, which is sufficient to separate the isotope peaks of a triply charged ion. The same instrument at m/z 4000 may deliver rather less, because resolving power is not constant across the mass range and the figure on a specification sheet is quoted at whichever mass flatters it.

The number that matters for identity work is whether the instrument can separate two species whose masses differ by the amount you care about. To distinguish a deamidated peptide from its parent at four thousand daltons requires separating peaks 0.98 apart, which is a resolving power of roughly four thousand — modest. To distinguish a glutamine-for-lysine substitution requires separating peaks 0.036 apart at the same mass, which is a resolving power above one hundred thousand. Those two requirements differ by a factor of twenty-five and both are described in the trade by the same phrase, high resolution.

The Journal’s practice, adopted after an exchange with a laboratory that pointed out we had been sloppy about it, is to state the discrimination rather than the specification: not “high-resolution MS”, but “sufficient to resolve a 0.98-dalton shift at the parent mass”. It is longer and it says something.1

Print the spectrum. It is the cheapest available improvement to identity reporting and it consists of reproducing a picture you already have.

The Journal’s request to laboratories, restated annually

Mass accuracy, and what a tolerance ought to be

Mass accuracy is the difference between the measured mass and the true mass, expressed in parts per million of the measured value. It depends on calibration, on the stability of the instrument’s electronics and temperature, on the number of ions arriving at the detector, and on whether an internal calibrant was co-analysed with the sample. It is not a fixed property of an instrument; it is a property of a measurement made on an instrument on a particular day.

Certificates in this market seldom state a tolerance at all. Where they do, the figure is usually expressed in daltons rather than parts per million and is generous: ±0.5 or ±1.0 dalton is common, which at incretin molecular weights corresponds to 120 to 240 parts per million and is achievable on almost any instrument sold in the last thirty years. A tolerance that no plausible measurement could fail is not an acceptance criterion. It is a formality.

What a meaningful criterion looks like is not mysterious. State the theoretical mass and its convention, state the observed mass, state the deviation in parts per million, and state the limit above which the result would have been reported as non-conforming. Four numbers, all of them already known to the analyst. The compendial framework for validating an analytical procedure asks for exactly this kind of specificity about what a test can discriminate, and the framework predates this market by decades.2

Common mass shifts, and the resolving power needed to see them at 4000 Da
ChangeMass shift (Da)Shift (ppm at 4000 Da)Resolving power required
Oxidation (one O added)+15.9953999~250
Deamidation of Asn or Gln+0.984246~4100
Disulphide formation−2.016504~2000
Pyroglutamate formation−18.0114503~220
TFA adduct+113.99328498~35
Gln replaced by Lys−0.0369~110000
Leu replaced by Ile0.0000not resolvable
L to D inversion0.0000not resolvable
Required resolving power estimated as the parent mass divided by the mass shift, which is the minimum needed to present the two species as separate peaks; in practice a factor of two above this figure is needed for reliable quantitation of the minor species.

The instrument classes, and what each can support

A single quadrupole mass filter provides unit resolution and mass accuracy of a few tenths of a dalton. It is entirely adequate to confirm that a sample is broadly the compound expected and to detect large modifications, and it is the analyser in most low-cost LC-MS systems. It cannot resolve an isotopic envelope at peptide molecular weights and therefore cannot assign charge from spacing.

Time-of-flight analysers separate ions by the time they take to traverse a flight tube. A linear tube gives modest resolving power; adding a reflectron and delayed extraction raises it into the tens of thousands, and modern quadrupole time-of-flight hybrids achieve low single-figure parts-per-million accuracy with routine calibration. Orbital trapping instruments measure the frequency of ion oscillation in an electrostatic field and convert it by Fourier transform, delivering resolving powers from sixty thousand to several hundred thousand and sub-part-per-million accuracy with internal calibration. Fourier-transform ion cyclotron resonance remains the highest-performing class and the least common outside academic facilities.

What this hierarchy means for a reader of certificates is that the instrument named on the document sets a ceiling on what the document can claim, independent of the laboratory’s competence. An unnamed instrument leaves that ceiling unknown, which is why the Journal now treats the absence of an instrument name as a material omission rather than a stylistic one.

Calibration, drift, and the internal standard

Every mass spectrometer is calibrated against a mixture of compounds of known exact mass, and every mass spectrometer drifts away from that calibration afterwards. The rate depends on the analyser type, on ambient temperature stability, and in trapping instruments on the number of ions in the trap: space-charge effects shift apparent masses in a manner that depends on how much sample was injected.

External calibration means the calibrant was run separately, before or after the samples. It is simple, it is what most routine work uses, and it is vulnerable to everything that happens between the calibration and the sample. Internal calibration means a compound of known mass was present in the same spectrum as the analyte, so the correction is applied to the measurement rather than to the instrument. Internal calibration is the reason sub-part-per-million figures are achievable at all, and it is the difference between a stated accuracy and a demonstrated one.

None of this is exotic or contested; it is ordinary laboratory practice, described in accreditation requirements as part of metrological traceability and in the pharmacopoeial chapters as part of system suitability.3 The reason it belongs in an article aimed at buyers is that it explains why two competent laboratories analysing the same vial on the same class of instrument can differ by tens of parts per million, and why the honest response to such a difference is to ask about calibration rather than about honesty.

1097795492190-1130.98 Da shift0.036 Da shift1000200030004000500060008000peptide molecular weight (Da)ppm accuracy required
Figure. Mass accuracy required to resolve two shifts, as a function of peptide molecular weight. The upper series is deamidation at 0.98 Da; the lower is the glutamine-for-lysine exchange at 0.036 Da. The horizontal scale is where routine instruments sit: a few parts per million for an orbital trap, several hundred for a linear time-of-flight.

The standing rule in this department

This publication applies one rule to every identity claim it reports, and it is worth stating in isolation because it governs the rest. A mass measurement supports a statement about composition. Only a fragmentation or mapping experiment supports a statement about sequence. Where a source says identity was confirmed, we report that a mass was measured, unless we have seen evidence of the second kind.

The rule has consequences we accept. It makes our coverage read as more sceptical than the underlying documents, because the documents claim more than they establish. It occasionally irritates laboratories which have in fact done sequence-level work and have simply not printed it, and the remedy there is a two-line email which we are glad to receive. And it means we cannot describe any research-grade vial in this market as sequence-confirmed, because on the evidence available to us almost none are.

What the rule is not is an accusation. Nothing in this article suggests that vendors are selling material other than what they label, and the Journal has no evidence of that in respect of any company it covers. The claim is narrower and, we think, harder to argue with: the documentation in general circulation does not have the discriminating power that the language on it implies, and the gap between the two is where every avoidable dispute in this market begins.

A short glossary, because the terms are used loosely

m/z — mass-to-charge ratio, the quantity a mass spectrometer actually measures. Monoisotopic mass — mass calculated using the lightest stable isotope of each element. Average mass — mass calculated using standard atomic weights. Nominal mass — the integer sum of integer isotope masses; adequate for small molecules, useless here.

Resolving power — m divided by peak width at half height; the ability to separate nearby masses. Mass accuracy — deviation of a measurement from the true value, in parts per million. Mass defect — the difference between an exact mass and its nominal value, and the property that makes near-isobars separable.

Adduct — an ion formed by association with something other than a proton, commonly sodium or potassium. Charge-state envelope — the family of differently charged ions from one compound. Isobaric — of identical mass at the achieved precision. Isomeric — of identical composition and different structure. b and y ions — the complementary fragment series produced by amide-bond cleavage.

Precision in these terms is not decoration. Several disputes this department has been asked to adjudicate turned out, on inspection, to be disagreements about whether the word mass meant monoisotopic or average.

Readers who take one thing from this piece should take the arithmetic. Isotope spacing is one over the charge. Deamidation is one dalton, oxidation is sixteen, and stereochemical inversion is nothing at all. A tolerance of ±1 dalton on a four-thousand-dalton peptide is two hundred and forty parts per million and excludes almost nothing worth excluding. Those four facts are sufficient to read most of the identity claims in circulation, and they fit on the back of an envelope.

References

  1. “Resolving power, mass accuracy and the limits of composition assignment in high-resolution mass spectrometry.” Analytical Chemistry. 2019;91(4):2410–2421.
  2. United States Pharmacopeia. General chapter ⟨1225⟩, Validation of Compendial Procedures. USP–NF.
  3. International Organization for Standardization. ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories. Geneva, 2017. Clauses 6.4 and 6.5 on equipment and metrological traceability.

Letters to the Editor

4 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

I have spent a week trying to reconcile a certificate’s stated mass of 4113.6 with a figure of 4111.1 I calculated from the sequence, and had convinced myself something was wrong with the vial. It was the isotope convention. Thank you, and also: how is this not stated on every certificate in existence?

A. Basaraba, Winnipeg, MB

The Journal replies

We wish we knew. It is the single most common source of spurious discrepancies reaching this desk, it costs nothing to state, and we have now asked all twenty companies in the dossier programme to add it. Three have.

You write that leucine and isoleucine cannot be distinguished by tandem mass spectrometry. That is too absolute. Side-chain fragmentation under high-energy conditions produces diagnostic w and d ions, and the discrimination has been demonstrated repeatedly.

P. McAlinden, Belfast

The Journal replies

Correct, and the text has been amended. The discrimination is achievable under specialised conditions and is not available in any routine service this market uses, which is what we should have written rather than the stronger claim.

I would add one omission to your six lines: the date and nature of the last calibration. A parts-per-million figure from an instrument last calibrated a fortnight ago is a different claim from one calibrated that morning with an internal standard.

D. Chukwuma, Onitsha

The Journal replies

Agreed, and it may be the best suggestion we have received on this subject. It is now a seventh line in the version of the list we send to suppliers, with the note that internal calibration should be stated where it was used.

Your article says a matching mass does not confirm a sequence, which is correct, and then rather implies that vendors are trading on the ambiguity. I run analytical services and I would put it differently: we report what we measured, in the words our clients ask for. If the Journal wants the word confirmed retired, write to the buyers, not to us.

M. Sandhu, Amritsar

The Journal replies

That is a fair reallocation of the criticism and we accept it. The word is chosen by whoever commissions the report, and laboratories are answering the question they were paid to answer. Our complaint is with the practice, not with the analysts, and the article should have located it more precisely.

Related coverage

Analytics

Check the vial, not the box

The interval between manufacture and analysis is the most under-read figure on the page, and the one most likely to matter by the time a vial is opened.

Wren Kealoha·3 Jul 2024·14 min