Why the theoretical mass on the certificate might be the wrong theoretical mass
A worked example: one peptide, five charge states, five mass-to-charge ratios, one molecular weight.
TheCompound Journal
Reporting on incretins, compounding & the peptide supply chain
Page 2 of 5 of this archive, newest first.
A worked example: one peptide, five charge states, five mass-to-charge ratios, one molecular weight.
Peptide bonds absorb strongly near 214 nm; aromatic side chains absorb near 280 nm. A method reading at 280 is blind to any fragment lacking an aromatic residue.
None of what a checkable identity statement requires is commercially sensitive, and all of it is known to whoever produced the document.
The two ionisation techniques in general use produce different charge distributions, different adducts and different failure modes. Certificates almost never say which was…
Where two methods disagree, the conservative convention is to report the lower figure. It is not universal, and whether a laboratory follows it belongs on the report.
Chromatographic software does not integrate every fluctuation in the baseline. It applies a threshold, and the threshold changes the reported purity by amounts that matter…
We set out the parameters a method disclosure would contain, all of which are known to whoever ran the sample and none of which is commercially sensitive.
An intact mass measurement establishes elemental composition, at best. The number of distinct sequences consistent with a given composition is astronomically large.
What the Journal asks for when it writes to a supplier about an identity claim, and how often it gets it.
A shallow gradient resolves impurities that a steep one runs into the parent peak. Both methods are legitimate; only one of them can see the small stuff.
We asked the four independent testing services what standards they run against and what suitability criteria they apply. The answers are printed.
Parts per million sound impressive until they are converted into daltons at the molecular weight of the thing being measured.
What would be needed to catch each of these, and what it would cost.
We set out the parameters a method disclosure would contain, all of which are known to whoever ran the sample and none of which is commercially sensitive.
Parts per million sound impressive until they are converted into daltons at the molecular weight of the thing being measured.
Stereochemical inversion changes the molecule, changes its biology, and changes its mass by exactly nothing.
Matching a retention time against a standard is evidence of consistency, not proof of identity. Two different species can elute at the same time on one method.
We work through a single chromatogram twice, under two integration conventions, and show where the difference comes from.
An assay against a quantitative reference standard measures how much is present. Area per cent measures what proportion of the visible material is the parent. Only the first…
Calibration drift is real, unremarkable, and the reason serious laboratories run internal standards.