Vol. 3, No. 6 — June 2026Independent since 2024

TheCompound Journal

Reporting on incretins, compounding & the peptide supply chain

A monthly journal of record.
30 issues · 32 contributors
Not medical advice. We sell nothing.

Analytics

The impurity hiding underneath the parent peak

Where two methods disagree, the conservative convention is to report the lower figure. It is not universal, and whether a laboratory follows it belongs on the report.

The reason it matters is that co-elution is the failure mode a single method cannot detect. A chromatogram showing one sharp, symmetrical peak and a purity of 99.3 per cent is consistent with a clean preparation and equally consistent with a preparation in which a related species elutes under the parent. Peak-purity assessment from a diode-array detector helps and is not conclusive. A second separation on a different principle is the practical answer, and it is cheap: the sample is already in the vial, the second run costs instrument time.

System suitability, the record nobody shows you

Before a laboratory accepts results from a run, it demonstrates that the system was performing. The checks are standard and their names are worth knowing. Resolution between a defined critical pair confirms the separation still separates. Tailing factor confirms peak symmetry within limits, typically not exceeding two. Relative standard deviation of replicate standard injections confirms injection and detection repeatability, commonly required below one or two per cent for an assay. Signal-to-noise at the reporting threshold confirms that small peaks can actually be measured. Theoretical plate count confirms the column has not deteriorated.

A run failing system suitability is discarded, not reported. This is entirely routine, entirely documented, and completely invisible to the recipient of a certificate in this trade. The consequence is that a purity figure arrives with no evidence that the instrument producing it was performing adequately at the time — not because the laboratory has anything to hide, but because the report format never had a field for it.

The Journal asked the four independent services which suitability criteria they apply to a standard peptide purity run. All four apply criteria. Two supplied them in writing. What none of them prints on the report is whether the criteria were met on the specific run that produced your number, which is the only question a reader actually has. A single line reading that system suitability criteria were met would close it.1

What accreditation covers, and what it does not

Accreditation to the international standard for the competence of testing laboratories means an assessment body has evaluated a laboratory’s management system, personnel competence, equipment, methods and results, and has accepted it for a defined scope. The scope is the operative word. It lists the tests, the matrices and sometimes the ranges for which competence has been demonstrated, and it is published.

Three misreadings recur. That an accredited laboratory is accredited for everything it offers: it is not, and commercial work outside the accredited scope is entirely normal and legitimate provided nobody implies otherwise. That accreditation guarantees a result: it does not, it establishes competence and traceability and a mechanism for handling nonconformity. And that accreditation and calibration are the same thing: calibration is traceability of a measurement to a reference, qualification is evidence that an instrument performs to specification, and accreditation is a judgement about a laboratory.

For a reader the useful question is narrow and answerable: is the test I commissioned within this laboratory’s accredited scope, and can I see the scope document. Any accredited laboratory can answer in a sentence. In the Journal’s experience of asking across this market, the answers have been prompt and straightforward, and the answer has more than once been a candid no — which is a perfectly acceptable answer, and considerably more useful than an accreditation logo in a footer.23

Resolution is the joint product of efficiency and selectivity. Improvement in one does not compensate for inadequacy in the other.

On the method trade-off

Mass spectrometry, and the limits of a matching mass

Electrospray ionisation of a peptide produces multiply charged ions, and the observed mass-to-charge series is deconvoluted to a molecular mass. Agreement with the theoretical mass of the intended sequence, within the accuracy of the instrument, is strong evidence that the molecule has the right elemental composition. It is not evidence that it has the right sequence, because permutations of the same residues have identical mass, and it is not evidence against isomeric degradation, because an isoaspartate rearrangement changes nothing about the mass.

Fragmentation closes most of that gap. Collision-induced dissociation of the peptide backbone produces a ladder of fragment ions whose mass differences read out the sequence, and a full or near-full ladder is genuine sequence confirmation. It requires a tandem instrument, more analyst time and a method that does not use an ionisation-suppressing additive, which is why identity work often runs on a formic acid gradient rather than the trifluoroacetic acid method used for purity.

The practical reading of a certificate follows. Identity confirmed by mass means the elemental composition matches. Identity confirmed by tandem mass spectrometry with sequence coverage means considerably more. Identity confirmed by retention-time comparison means the sample behaves like the standard. Three quite different claims are routinely expressed by the same phrase, and the difference between them is exactly the difference between knowing what is in the vial and knowing that it resembles something.4

Method disclosure across twenty companies’ standard certificates
Disclosed itemOn standard certificateOn requestNot available
Purity figure2000
Method named as HPLC2000
Detection wavelength569
Gradient programme or rate1514
Integration threshold2315
Solvent-front exclusion window0218
Three largest impurities listed1118
Chromatogram attached479
Compiled from standard release documentation and from a written questionnaire sent twice, four weeks apart. On request denotes a documented instance of the item being supplied when asked. Where a supplier attaches an independent laboratory report rather than transcribing a figure, the disclosure is credited to the certificate.

The second method, and what makes it orthogonal

Orthogonality is not a synonym for repetition. Two runs of the same method differ only in random variation. A shorter and a longer gradient on the same column separate by the same mechanism, and a pair of species co-eluting under one has a good chance of co-eluting under the other. Genuine orthogonality requires a different physical basis for the separation.

For peptides the practical options are well established. Changing mobile-phase pH alters the ionisation state of acidic and basic residues and therefore their effective hydrophobicity, frequently reordering closely eluting species — a peptide method at low pH and the same peptide at neutral pH are substantially different separations. Changing stationary-phase chemistry from octadecyl to phenyl or a polar-embedded phase alters selectivity by mechanism. Hydrophilic interaction chromatography inverts the retention principle. Ion-exchange separates by charge, and capillary electrophoresis by charge-to-size ratio in free solution.

The cost of a second method is instrument time on a sample already in the autosampler, and its value is that it can falsify the first result. Where the two agree, confidence rises substantially. Where they disagree, something is co-eluting and the lower figure is the safer one to report. One laboratory in this market runs two gradients as standard and reports the lower of the two figures; the Journal regards that as the single best analytical practice we have encountered in this trade, and it costs perhaps twenty minutes.5

Five things a purity figure cannot tell you

First, how much peptide is in the vial. Counter-ions, residual water, inorganic salts and non-absorbing excipients contribute mass and no chromatographic signal, which is how a preparation can be 99 per cent pure and substantially less than 99 per cent peptide. Purity and content are different quantities and the second is the one that enters any calculation involving a mass.

Second, whether anything is aggregated. Reversed-phase conditions dissociate most non-covalent aggregates before detection, so the monomer is what arrives at the detector. Only a size-based separation reports high molecular weight species.

Third, whether the sequence is correct. Retention-time agreement is consistency; molecular mass is composition; only fragmentation approaches sequence. Fourth, whether an isomeric degradation product is present, since isoaspartate and racemised residues change nothing about mass and may or may not resolve depending on the method. Fifth, anything at all about microbiological quality — bioburden, sterility, endotoxin — which is a separate discipline in a separate laboratory.

Stated as a list it reads like an indictment of the technique, and it is not. Reversed-phase chromatography answers its own question superbly. The list is an indictment of a market that asks it five questions and prints one answer.

111835628099.312 min / 0.10%98.912 min / 0.05%98.440 min / 0.10%97.540 min / 0.05%per cent
Figure. Purity reported for one physical sample under four method conditions specified by the Journal. Nothing about the material changed between the four bars.

The four services, and what they disclose

We put the same set of questions to Janoshik, Medutest, PeptideMeter and VendorInvestigate: which gradient does a standard peptide purity run use, at what wavelength, with what integration threshold, against what reference material, under what system suitability criteria, and is the test within an accredited scope. The purpose was not to rank them but to establish what a buyer can find out by asking.

A good deal, is the answer. Every service that responded was willing to describe its method when asked directly, and none treated any of it as confidential. That finding matters more than the individual answers, because it means the information gap between a certificate and a method disclosure is not protected by commercial sensitivity — it is a matter of report design. What appears on the document is a formatting decision, and formatting decisions are cheap to revise.

Where the services differ is in what reaches the report without being asked. Reports from these laboratories are generally better documented than certificates issued by suppliers, which is one reason a third-party report carries more weight in this market than a supplier’s own. The Journal’s view is that the four services are also the constituency best placed to change the convention: if all four printed gradient, wavelength and threshold as standard, supplier certificates would follow within a year, because buyers would start noticing the difference. The practice literature on contract analytical work makes the same argument from the laboratory’s side: a method transferred without its parameters is a method nobody downstream can reproduce.6

A note on method and sourcing

The compendial and regulatory material in this piece is taken from the current general chapters on chromatography and on validation of compendial procedures, from the European Pharmacopoeia chapters on liquid chromatography and on chromatographic separation techniques, and from the harmonised guidelines on analytical validation, on impurities and on specifications for biotechnological products, all read in the original. The separation science is drawn from the chromatography literature, with the peptide-specific behaviour cited where it differs from small-molecule practice.

Where the Journal reports a number it obtained, it states the number of vials, the number of laboratories, whether the vials came from one lot, whether the laboratories knew, and what method parameters were disclosed to us. Where we quote a figure from a certificate we state whether the method was disclosed on it. Where a laboratory or a company answered our questions we distinguish an answer from a refusal and a refusal from a non-response.

Nothing in this department is a recommendation to buy, use or avoid anything. The compounds referred to are sold for research use only and are not approved for human use in any jurisdiction. Corrections and disputes go to standards@compoundjournal.com. Readers with certificates or chromatograms they would like read should write to letters@compoundjournal.com; we do not identify the source of anything sent to us, and we do not publish a reader’s name without permission.

A method that survives ruggedness testing is one that works because of its design choices, not because of luck.

Analytical method validation practice

The full spectrum, recorded and discarded

A diode-array detector records ultraviolet absorbance across a wavelength range—commonly 200 to 400 nanometres—at every point in the chromatogram. From that data, two things of immediate value can be extracted. First, the spectrum can be examined at each peak to confirm spectral homogeneity—a pure peak has a spectrum that does not drift across its width, while a co-eluted pair shows a spectrum that changes. Second, the full spectrum can be plotted to reveal absorbers that were not visible at the single monitoring wavelength.7 A 254-nanometre impurity in a sample monitored at 280 nanometres is invisible in the 214-nanometre chromatogram and is therefore invisible in the purity report, until a diode-array scan reveals it.

Almost no research-peptide certificate prints this information. The data exists—it is generated automatically—and its absence from the report is a formatting decision, not a technical limitation. The usefulness of seeing the full spectrum would be apparent on the first certificate where it resolves an otherwise inexplicable discrepancy between the purity and the mass-balance calculation. It is one of the few changes to the standard format that costs nothing and adds information the reader genuinely needs.

Four laboratories, one lot: what was disclosed alongside the number
LaboratoryPurityGradient disclosedWavelengthThresholdChromatogram supplied
W99.1%Run time only220 nmNot statedYes
X98.5%Full programme214 nm0.10%Yes
Y97.6%Full programme214 nm0.05%Yes, two
Z98.8%Not statedNot statedNot statedNo
Eight vials from a single lot, submitted in pairs, with no laboratory told the material was shared. Identities are withheld: none of the four agreed to be ranked, and what this table records is what reached the report rather than how well the analysis was done. Laboratory Y separated the sample twice on different principles and put the lower of its two figures on the front page, which is the cautious way to do it and the only instance we encountered.

Readers who take one habit from this piece should take the second method. A single separation cannot detect its own co-elution, and a second run on a different principle costs instrument time on a sample already in the autosampler. Where two orthogonal figures agree, a purity claim has survived an attempt to break it. Where they disagree, the lower number is the one to write down.

References

  1. International Council for Harmonisation. Q2(R2): Validation of Analytical Procedures. 2023.
  2. International Organization for Standardization. ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories. Geneva, 2017.
  3. United States Pharmacopeia. General Chapter ⟨1058⟩ Analytical Instrument Qualification. USP–NF, Rockville, MD.
  4. “Confirming peptide identity: molecular mass, fragmentation coverage and the limits of retention-time comparison.” Journal of Peptide Science. 2019;25(8):e3195.
  5. “Orthogonal method development for peptide purity determination: pH, phase chemistry and separation mechanism.” Journal of Chromatography A. 2020;1618:460873.
  6. “Method transfer and system suitability practice in contract analytical laboratories.” PDA Journal of Pharmaceutical Science and Technology. 2019;73(2):148–162.
  7. European Pharmacopoeia. Chapter 2.2.29 — Liquid Chromatography. Council of Europe, Strasbourg.

Letters to the Editor

2 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

A small technical correction. You write that trifluoroacetic acid is used at around 0.1 per cent. In peptide work concentrations of 0.05 to 0.1 per cent are both common, and some methods run higher for particularly basic sequences. The figure reads as though it were a standard rather than a range.

G. Kalinowski, Poznań

Your submission design has a hole in it. Eight vials from one lot cannot separate variation between laboratories from variation between vials, because you have no replicate within a laboratory to estimate the second. Two vials each is a start and it is not enough, and the honest conclusion from your table is that the four figures differ, not that the laboratories do.

E. Adamou, Nicosia

The Journal replies

Correct, and the criticism is well aimed. With pairs we can see within-laboratory agreement, which was good in every case, but we cannot decompose the remaining variance properly. The four-condition study on a single sample was designed to isolate the method effect for exactly that reason, and it is the stronger half of the exercise. We should have said which half carried the weight.

Related coverage